Evidence map›Paper›PMID 36269884›Full record

ArticleTexas Heart Institute journal2022

Comparison of MANTA vs ProGlide Vascular Closure Device and 30-Day Outcomes in Transfemoral Transcatheter Aortic Valve Implantation.

Christos Eftychiou, Nicolaos Eteocleous, Andreas Mitsis, Ioannis Zittis, Kyriacos Papadopoulos, Andria Petrou, Maria Adamou, Christos Kounnos, Christos Bekos, Panayiotis Avraamides

Open access · diamondAbstract read
In one paragraph

Article in Texas Heart Institute journal, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
1.0field-weighted citation impact, top 23% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 6 citations in OpenAlex.

  1. Collagen-Based Vascular Closure Device Multicenter Italian Experience in Endovascular Aortic Aneurysm Repair Compared With Suture-Mediated Closure Vascular Device.Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2026
    Observational
  2. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors at 1 institution in 1 country.

Christos EftychiouCardiology Department, Nicosia General Hospital, Nicosia, Cyprus.
Nicolaos EteocleousCardiology Department, Nicosia General Hospital, Nicosia, Cyprus.
Andreas MitsisCardiology Department, Nicosia General Hospital, Nicosia, Cyprus.
Ioannis ZittisCardiology Department, Nicosia General Hospital, Nicosia, Cyprus.
Kyriacos PapadopoulosCardiology Department, Nicosia General Hospital, Nicosia, Cyprus.
Andria PetrouAnesthesiology Department, Nicosia General Hospital, Nicosia, Cyprus.
Maria AdamouBlood Bank Department, Nicosia General Hospital, Nicosia, Cyprus.
Christos KounnosVascular Surgery Department, Nicosia General Hospital, Nicosia, Cyprus.
Christos BekosVascular Surgery Department, Nicosia General Hospital, Nicosia, Cyprus.
Panayiotis AvraamidesCardiology Department, Nicosia General Hospital, Nicosia, Cyprus.
Nicosia General Hospital · CY

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundVascular complications (VCs) after transfemoral transcatheter aortic valve implantation (TAVI) have an increased mortality risk, and vascular closure device (VCD) use is mandatory. The percutaneous MANTA VCD (Teleflex) is a novel collagen-based technology for closure of large-bore arteriotomies. We compared the MANTA VCD with the suture-based ProGlide VCD (Abbott Vascular).

methodsA retrospective review was performed on all consecutive patients who underwent transfemoral TAVI in our center from January 1, 2015, to February 28, 2021, and 30-day outcomes were recorded. Major adverse cardiovascular events (MACE) were cardiac death, disabling stroke, and/or major VCs. Access site-related VCs were VCs related to the access site vessel from which the transcatheter valve was introduced and advanced.

resultsThe MANTA VCD was used in 99 patients and the ProGlide in 224. There was 4.0% MACE in the MANTA group and 4.9% in the ProGlide group (P = .999). Overall VCs were 10.1% vs 7.6%, major VCs were 3.0% vs 2.2%, and minor VCs 7.1% vs 5.4%(P = .753). Access site-related VCs were 5.1% vs 5.8% in the (P = .999), and periprocedural vascular surgical intervention was needed in 6.1% of the MANTA group vs 2.2% of the ProGlide group (P = .099).

conclusionThere was no significant difference in MACE, mortality, cardiovascular mortality, VCs, access site-related VCs, periprocedural vascular surgical interventions, bleeding, or transfusion rate between the 2 groups. The MANTA VCD group had more periprocedural vascular surgical interventions which did not reach statistical significance.

Indexed as

Aortic Valve StenosisTranscatheter Aortic Valve ReplacementVascular Closure DevicesAortic ValveCollagenFemoral ArteryHumansTreatment OutcomeCollagenaortic valve diseaseTAVIvascular closure devicevascular complications

Identifiers

PMID36269884
PMCPMC9632403
OpenAlexW4307050798

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked