Evidence map›Paper›PMID 36157896›Full record

ArticleMedicine in novel technology and devices2022

Computational fluid dynamics analysis and experimental hemolytic performance of three clinical centrifugal blood pumps: Revolution, Rotaflow and CentriMag.

Dong Han, Joshua L Leibowitz, Lu Han, Shigang Wang, Ge He, Bartley P Griffith, Zhongjun J Wu

Open access · goldAbstract read
In one paragraph

Article in Medicine in novel technology and devices, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 14 papers.

0numbers the graph read from it
0cells of the map it votes in
14citing papers in PubMed
3.1field-weighted citation impact, top 8% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

14 citing papers in PubMed, 41 citations in OpenAlex.

  1. Review
  2. Article
  3. Computational and Experimental Assessment of Shear-Induced Blood Trauma by HeartMate II, HeartMate 3, and BrioVAD.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
    Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 1 institution in 1 country.

Dong HanDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, USA.
Joshua L LeibowitzDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, USA.
Lu HanDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, USA.
Shigang WangDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, USA.
Ge HeDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, USA.
Bartley P GriffithDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, USA.
Zhongjun J WuDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, USA.
University of Maryland, Baltimore · US

Funding

First-in-Human Use of an Artificial Lung for Ambulatory Respiratory SupportR01HL118372 · NHLBI · UNIVERSITY OF MARYLAND BALTIMORE · PI GRIFFITH, BARTLEY P, WU, ZHONGJUN JON · 2013 to 2023
$7.9M
NRSA Training CoreTL1TR003100 · NCATS · JOHNS HOPKINS UNIVERSITY · PI SAHA, SOMNATH · 2019 to 2023
$4.5M
Development and Pre-Clinical Trial Qualification of a Pediatric Pump-Lung System for ChildrenR01HL141817 · NHLBI · UNIVERSITY OF MARYLAND BALTIMORE · PI GRIFFITH, BARTLEY P, WU, ZHONGJUN JON · 2018 to 2021
$3.1M
Supplement: Hemolysis Prediction Software DevelopmentR01HL131750 · NHLBI · LEHIGH UNIVERSITY · PI LIU, YALING · 2017 to 2021
$2.2M
Shear-Induced Hemostatic Dysfunction and Bleeding in CF-VAD PatientsR01HL124170 · NHLBI · UNIVERSITY OF MARYLAND BALTIMORE · PI GRIFFITH, BARTLEY P, WU, ZHONGJUN JON · 2015 to 2018
$2.0M
NCATS NIH HHS TL1 TR003100NHLBI NIH HHS R01 HL118372NHLBI NIH HHS R01 HL124170NHLBI NIH HHS R01 HL131750NHLBI NIH HHS R01 HL141817
6 · The paper itself

Abstract

Centrifugal blood pumps have become popular for adult extracorporeal membrane oxygenation (ECMO) due to their superior blood handling and reduced thrombosis risk featured by their secondary flow paths that avoid stagnant areas. However, the high rotational speed within a centrifugal blood pump can introduce high shear stress, causing a significant shear-induced hemolysis rate. The Revolution pump, the Rotaflow pump, and the CentriMag pump are three of the leading centrifugal blood pumps on the market. Although many experimental and computational studies have focused on evaluating the hydraulic and hemolytic performances of the Rotaflow and CentriMag pumps, there are few on the Revolution pump. Furthermore, a thorough direct comparison of these three pumps' flow characteristics and hemolysis is not available. In this study, we conducted a computational and experimental analysis to compare the hemolytic performances of the Revolution, Rotaflow, and CentriMag pumps operating under a clinically relevant condition, i.e., the blood flow rate of 5 L/min and pump pressure head of 350 mmHg, for adult ECMO support.

Indexed as

Centrifugal blood pumpsCFDECMO supportHemolysis

Identifiers

PMID36157896
PMCPMC9497451
OpenAlexW4283156302

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.