Iron-based coupling media for MRI-guided ultrasound surgery.
Article in Medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed, 4 citations in OpenAlex.
- Characterization of a Next-Generation Iron Oxide Coupling Medium for Transcranial Magnetic Resonance-Guided Focused Ultrasound.Ultrasound in medicine & biology · 2026Article
- A retraced spiral strategy with semi-automatic deblurring for volumetric thermometry.Magnetic resonance in medicine · 2025Article
- Probability of Cavitation in a Custom Iron-Based Coupling Medium for Transcranial Magnetic Resonance-Guided Focused Ultrasound Procedures.Ultrasound in medicine & biology · 2023Article
- Dark band artifact in transcranial MR-guided focused ultrasound: Mechanism and mitigation with passive crossed wire antennas.Magnetic resonance imaging · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 4 institutions in 1 country.
Funding
Abstract
purposeIn this study, we examine the effects of a recently developed, iron-based coupling medium (IBCM) on guidance magnetic resonance (MR) scans during transcranial, magnetic-resonance-guided, focused ultrasound surgery (tMRgFUS) procedures. More specifically, this study tests the hypotheses that the use of the IBCM will (a) not adversely affect image quality, (b) remove aliasing from small field-of-view scans, and (c) decouple image quality from the motion state of the coupling fluid.
methodsAn IBCM, whose chemical synthesis and characterization are reported elsewhere, was used as a coupling medium during tMRgFUS procedures on gel phantoms. Guidance magnetization-prepared rapid-gradient-echo (MP-RAGE), TSE, and GRE scans were acquired with fields of view of 28 and 18 cm. Experiments were repeated with the IBCM in several distinct flow states. GRE scans were used to estimate temperature time courses as a gel target was insonated. IBCM performance was measured by computing (i) the root mean square difference (RMSD) of TSE and GRE pixel values acquired using water and the IBCM, relative to the use of water; (ii) through-time temperature uncertainty for GRE scans; and (iii) Bland-Altman analysis of the temperature time courses. Finally, guidance TSE and GRE scans of a human volunteer were acquired during a separate sham tMRgFUS procedure. As a control, all experiments were repeated using a water coupling medium.
resultsUse of the IBCM reduced RMSD in TSE scans by a factor of 4 or more for all fields of view and nonstationary motion states, but did not reduce RMSD estimates in MP-RAGE scans. With the coupling media in a stationary state, the IBCM altered estimates of temperature uncertainty relative to the use of water by less than 0.2°C. However, with a high flow state, the IBCM reduced temperature uncertainties by the statistically significant amounts (at the 0.01 level) of 0.5°C (28 cm field of view) and 5°C (18 cm field of view). Bland-Altman analyses found a 0.1°C ± 0.5°C difference between temperature estimates acquired using water and the IBCM as coupling media. Finally, scans of a human volunteer using the IBCM indicate more conspicuous grey/white matter contrast, a reduction in aliasing, and a less than 0.2°C change in temperature uncertainty.
conclusionsThe use of an IBCM during tMRgFUS procedures does not adversely affect image quality for TSE and GRE scans, can decouple image quality from the motion state of the coupling fluid, and can remove aliasing from scans where the field of view is set to be much smaller than the spatial extent of the coupling fluid.