ArticleBMC medical informatics and decision making2022
Concordance of three point of care testing devices with clinical chemistry laboratory standard assays and patient-reported outcomes of blood sampling methods.
Article in BMC medical informatics and decision making, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed, 1 synthesis or guideline pooled it, 9 citations in OpenAlex.
- Pooled it
- Effectiveness of multi-stage diabetes screening in the dental care setting: a scoping literature review.BMC oral health · 2026Article
- HbA1c measurement using capillary blood in epidemiological studies: a scoping review of technologies and approaches.BMC endocrine disorders · 2026Article
- A Digital Diabetes Prevention Program for Hispanic Adolescents (Fit24+): Protocol for a Feasibility Randomized Controlled Trial.JMIR research protocols · 2026Article
- Role of Point-of-Care Diagnostics in Lower- and Middle-Income Countries and Austere Environments.Diagnostics (Basel, Switzerland) · 2023Review
- Lab on a Chip Device for Diagnostic Evaluation and Management in Chronic Renal Disease: A Change Promoting Approach in the Patients' Follow Up.Biosensors · 2023Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors at 4 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPoint of care testing (POCT) devices have been developed to facilitate immediate results with the potential to aid screening for new disease and enable patients to self-monitor their disease. Non-communicable diseases (NCDs) are the major cause of mortality globally and are increasing in prevalence as the population ages. Allied health care professionals (AHPs) are skilled in undertaking risk assessment and delivering preventative advice, providing opportunities to access large proportions of the population who may not visit their doctor, within non-traditional community settings. There is evidence of high levels of support from public, patients and health professionals for engaging AHPs in risk-targeted early case detection of certain NCDs. Thus, POCT devices offer a potential alternative to traditional venous blood collection, as novel care pathways for increasing early case detection and access to preventative care. The objectives of this study were to: (i) determine the concordance of the specific POCT devices with laboratory-based standard assays employed within clinical biochemistry laboratories. (ii) compare the sampling experience of both methods via patient-reported experiences.
methodsA prospective, two-centre study was undertaken involving 158 participants who provided informed consent. Venous blood was collected for traditional assays of HbA1c, creatinine/ estimated Glomerular-Filtration-Rate (eGFR) and vitamin-D. Capillary blood was collected by finger prick test and also assayed for the same biochemical indices (Nova StatSensor (creatinine/eGFR); Siemens DCA-Vantage (HbA1C); CityAssays (vitamin-D)). All users were provided with device training. Participants reported any discomfort experienced by each simultaneously applied method (randomised in order) via a 100 mm Visual-Analogue-Scale.
resultsResults for each POCT device and the laboratory standard were analysed by Bland-Altman plots to determine assay concordance. POCT devices demonstrated good concordance with laboratory testing, with at least 95% of all samples being within two standard deviations, for each of the devices tested. The majority of participants reported less discomfort with POCT than venepuncture, with the average reported discomfort being 17/100 mm less for POCT compared to venous blood sample collection on the visual analogue scale.
conclusionsThe POCT devices demonstrated acceptable concordance with laboratory-based assays, and patients reported lower levels of discomfort compared to traditional means of blood collection. This study demonstrates the potential of using these devices as acceptable methods for opportunistic testing of "at-risk" individuals within non-traditional community care settings.