Locking mechanism failure between tibial baseplate and polyethylene insert in cruciate retaining total knee arthroplasty.

Anirudh Sharma, Vijay Killampalli, Arpit Patel

Open access · bronzeAbstract readCase Reports
In one paragraph

Article in Journal of clinical orthopaedics and trauma, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.4field-weighted citation impact, top 36% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 2 citations in OpenAlex.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 1 institution in 1 country.

Anirudh SharmaHinchingbrooke Hospital, North West Anglia NHS Foundation Trust, Huntingdon, Cambridgeshire, PE29 6NT, United Kingdom.
Vijay KillampalliHinchingbrooke Hospital, North West Anglia NHS Foundation Trust, Huntingdon, Cambridgeshire, PE29 6NT, United Kingdom.
Arpit PatelHinchingbrooke Hospital, North West Anglia NHS Foundation Trust, Huntingdon, Cambridgeshire, PE29 6NT, United Kingdom.
Hinchingbrooke Hospital · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

We describe a case of total knee arthroplasty performed with one type of cruciate retaining knee prosthesis (Vanguard CR, Zimmer Biomet), wherein failure of locking mechanism occurred between the tibial baseplate and the polyethylene insert 10 months after initial surgery. While there are very few case reports in literature describing such locking mechanism failure, all of these have been in constrained and posterior stabilized knees. This case, to the best of our knowledge is the first such case seen with a cruciate retaining knee design. A 73-year old lady, having undergone left total knee arthroplasty with the above-mentioned prosthesis, presented 10 months after the index surgery with a 3-week history of worsening knee pain and medially protruding hardware. She mentioned an episode of minor twisting of her knee after which she developed pain and swelling over the medial aspect, and was found to have the locking bar backed out on X-rays. She underwent revision surgery, wherein the locking bar was found to have buttonholed through the medial capsule. She underwent polyethylene liner exchange, with a new locking bar. There was no evidence of infection. Her symptoms fully settled following revision surgery.

Indexed as

Cruciate retaining kneeKnee arthroplasty failureVanguard knee