Pragmatic Approaches to Reducing Radiation Dose in Brain Computed Tomography Scan using Scan Parameter Modification.

Mohammad Reza Choopani, Iraj Abedi, Fatemeh Dalvand

Abstract read
In one paragraph

Article in Journal of medical signals and sensors. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Mohammad Reza ChoopaniDepartment of Radiology, Isfahan University of Medical Sciences, Isfahan, Iran.
Iraj AbediDepartment of Medical Physics, Isfahan University of Medical Sciences, Isfahan, Iran.
Fatemeh DalvandDepartment of Medical Radiation Engineering, Shahid Beheshti University, Tehran, Iran.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: High radiation dose of patients has become a concern in the computed tomography (CT) examinations. The aim of this study is to guide the radiology technician in modifying or optimizing the underlying parameters of the CT scan to reduce the patient radiation dose and produce an acceptable image quality for diagnosis. Methods: The body mass measurement device phantom was repeatedly scanned by changing the scan parameters. To analyze the image quality, software-based and observer-based evaluations were employed. To study the effect of scan parameters such as slice thickness and reconstruction filter on image quality and radiation dose, the structural equation modeling was used. Results: By changing the reconstruction filter from standard to soft and slice thickness from 2.5 mm to 5 mm, low-contrast resolution did not change significantly. In addition, by increasing the slice thickness and changing the reconstruction filter, the spatial resolution at different radiation conditions did not significantly differ from the standard irradiation conditions ( Conclusion: In this study, it was shown that in the brain CT scan imaging, the radiation dose was reduced by 30%-50% by increasing the slice thickness or changing the reconstruction filter. It is necessary to adjust the CT scan protocols according to clinical requirements or the special conditions of some patients while maintaining acceptable image quality.

Indexed as

Multidetector computed tomography scansradiation dosescan parameter modificationssoftware- and observer-based evaluatiosn