Characterization of left atrial assist device implantation: Early results of ex vivo anatomical assessment.
Article in Artificial organs, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
backgroundThe left atrial assist device (LAAD) is a novel pump that was developed specifically for the treatment of heart failure with preserved ejection fraction. The device is surgically implanted in the mitral position. This study aimed to characterize the various device-fitting configurations in the mitral annular position.
methodsRapidly prototyped LAAD models (n = 5) were fabricated with five different driveline configurations: (A) annulus level/intra-cuff running; (B) supra-cuff/below coronary sinus (CS); (C) infra-cuff; (D) supra-annulus/supra-CS; (E) left ventricular free wall level. The 3D-printed models were implanted in extracted fresh porcine hearts (80-100 kg, adult, healthy porcine) and the proximity of anatomical structures between the driveline and CS and coronary artery (CA) were measured.
resultsAll five device configurations were evaluated for fitting. For the purpose of preventing blood clot formation around the driveline, the mitral annulus (MA) as a driveline pass-way (configuration A) has been considered advantageous with the current device, in that the driveline exposure to blood has been avoided. The CS does not exist at exactly the same level as the MA, and there is less risk of injuring it than using the left atrial free wall. However, there is an inevitable risk of damaging the CA, so careful visual inspection before inserting the driveline is needed.
conclusionsSeveral options of driveline exteriorization were demonstrated, and the safety of each configuration was evaluated. Using the MA as a pathway for the driveline exit is considered to be a reasonable and safe method.