ArticleMedical physics2022

Task-based validation and application of a scanner-specific CT simulator using an anthropomorphic phantom.

Sachin S Shankar, Nicholas Felice, Eric A Hoffman, Jarron Atha, Jessica C Sieren, Ehsan Samei, Ehsan Abadi

Open access · greenAbstract read
In one paragraph

Article in Medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 16 papers.

0numbers the graph read from it
0cells of the map it votes in
16citing papers in PubMed
3.2field-weighted citation impact, top 9% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

16 citing papers in PubMed, 13 citations in OpenAlex.

  1. Article
  2. Article
  3. Observational
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Review
  10. Article
  11. Article
  12. Article
  13. Quantitative accuracy of lung function measurement using parametric response mapping: A virtual imaging study.Proceedings of SPIE--the International Society for Optical Engineering · 2024
    Article
  14. Article
  15. Development and Application of a Virtual Imaging Trial framework for Airway Quantifications via CT.Proceedings of SPIE--the International Society for Optical Engineering · 2023
    Article
  16. A systematic assessment of photon-counting CT for bone mineral density and microarchitecture quantifications.Proceedings of SPIE--the International Society for Optical Engineering · 2023
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 3 institutions in 1 country.

Sachin S ShankarCenter for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University, Durham, North Carolina, USA.
Nicholas FeliceCenter for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University, Durham, North Carolina, USA.
Eric A HoffmanDepartment of Radiology, University of Iowa, Iowa City, Iowa, USA.
Jarron AthaDepartment of Radiology, University of Iowa, Iowa City, Iowa, USA.
Jessica C SierenDepartment of Radiology, University of Iowa, Iowa City, Iowa, USA.
Ehsan SameiCenter for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University, Durham, North Carolina, USA.
Ehsan AbadiCenter for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University, Durham, North Carolina, USA.
Advanced Imaging Research (United States) · USUniversity of Iowa · USDuke University · US

Funding

Pulmonary Toxicology Facility CoreP30ES005605 · NIEHS · UNIVERSITY OF IOWA · PI Brandi Janssen · 1990 to 2026
$40.5M
TR&D Project 3: Virtual ReadersP41EB028744 · NIBIB · DUKE UNIVERSITY · PI Ehsan Samei · 2021 to 2026
$7.8M
Accuracy and Precision in CT Quantification of COPD Through Virtual Imaging TrialsR01HL155293 · NHLBI · DUKE UNIVERSITY · PI ABADI, EHSAN · 2021 to 2025
$2.2M
NHLBI NIH HHS R01 HL155293NIBIB NIH HHS P41 EB028744NIEHS NIH HHS P30 ES005605NIH HHS P41EB028744NIH HHS R01HL155293
6 · The paper itself

Abstract

backgroundQuantitative analysis of computed tomography (CT) images traditionally utilizes real patient data that can pose challenges with replicability, efficiency, and radiation exposure. Instead, virtual imaging trials (VITs) can overcome these hurdles through computer simulations of models of patients and imaging systems. DukeSim is a scanner-specific CT imaging simulator that has previously been validated with simple cylindrical phantoms, but not with anthropomorphic conditions and clinically relevant measurements. PURPOSE: To validate a scanner-specific CT simulator (DukeSim) for the assessment of lung imaging biomarkers under clinically relevant conditions across multiple scanners using an anthropomorphic chest phantom, and to demonstrate the utility of virtual trials by studying the effects or radiation dose and reconstruction kernels on the lung imaging quantifications.

methodsAn anthropomorphic chest phantom with customized tube inserts was imaged with two commercial scanners (Siemens Force and Siemens Flash) at 28 dose and reconstruction conditions. A computational version of the chest phantom was used with a scanner-specific CT simulator (DukeSim) to simulate virtual images corresponding to the settings of the real acquisitions. Lung imaging biomarkers were computed from both real and simulated CT images and quantitatively compared across all imaging conditions. The VIT framework was further utilized to investigate the effects of radiation dose (20-300 mAs) and reconstruction settings (Qr32f, Qr40f, and Qr69f reconstruction kernels using ADMIRE strength 3) on the accuracy of lung imaging biomarkers, compared against the ground-truth values modeled in the computational chest phantom.

resultsThe simulated CT images matched closely the real images for both scanners and all imaging conditions qualitatively and quantitatively, with the average biomarker percent error of 3.51% (range 0.002%-18.91%). The VIT study showed that sharper reconstruction kernels had lower accuracy with errors in mean lung HU of 84-94 HU, lung volume of 797-3785 cm

conclusionWe comprehensively evaluated the realism of DukeSim in an anthropomorphic setup across a diverse range of imaging conditions. This study paves the way toward utilizing VITs more reliably for conducting medical imaging experiments that are not practical using actual patient images.

Indexed as

Tomography, X-Ray ComputedComputer SimulationHumansPhantoms, ImagingRadiation DosageTomography Scanners, X-Ray ComputedCTCT simulatorDukeSimimage quality validationlung quantificationsvirtual imaging trial

Identifiers

PMID36097259
PMCPMC9792443
OpenAlexW4295507183

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.