Online geometry calibration for retrofit computed tomography from a mouse rotation system and a small-animal imager.
Article in Medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Investigating Image Distortion and Its Impact on Material Quantification in Photon-counting Computed Tomography.Journal of medical physicsArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
backgroundComputed tomography (CT) generates a three-dimensional rendering that can be used to interrogate a given region or desired structure from any orientation. However, in preclinical research, its deployment remains limited due to relatively high upfront costs. Existing integrated imaging systems that provide merged planar X-ray also dwarfs CT popularity in small laboratories due to their added versatility. PURPOSE: In this paper, we sought to generate CT-like data using an existing small-animal X-ray imager with a specialized specimen rotation system, or MiSpinner. This setup conforms to the cone-beam CT (CBCT) geometry, which demands high spatial calibration accuracy. Therefore, a simple but robust geometry calibration algorithm is necessary to ensure that the entire imaging system works properly and accurately.
methodsBecause the rotation system is not permanently affixed, we propose a structure tensor-based two-step online (ST-TSO) geometry calibration algorithm. Specifically, two datasets are needed, namely, calibration and actual measurements. A calibration measurement detects the background of the system forward X-ray projections. A study on the background image reveals the characteristics of the X-ray photon distribution, and thus, provides a reliable estimate of the imaging geometry origin. Actual measurements consisted of an X-ray of the intended object, including possible geometry errors. A comprehensive image processing technique helps to detect spatial misalignment information. Accordingly, the first processing step employs a modified projection matrix-based calibration algorithm to estimate the relevant geometric parameters. Predicted parameters are then fine-tuned in a second processing step by an iterative strategy based on the symmetry property of the sum of projections. Virtual projections calculated from the parameters after two-step processing compensate for the scanning errors and are used for CT reconstruction. Experiments on phantom and mouse imaging data were performed to validate the calibration algorithm.
resultsOnce system correction was conducted, CBCT of a CT bar phantom and a cohort of euthanized mice were analyzed. No obvious structure error or spatial artifacts were observed, validating the accuracy of the proposed geometry calibration method. Digital phantom simulation indicated that compared with the preset spatial values, errors in the final estimated parameters could be reduced to 0.05° difference in dominant angle and 0.5-pixel difference in dominant axis bias. The in-plane resolution view of the CT-bar phantom revealed that the resolution approaches 150
conclusionsA constrained two-step online geometry calibration algorithm has been developed to calibrate an integrated X-ray imaging system, defined by a first-step analytical estimation and a second-step iterative fine-tuning. Test results have validated its accuracy in system correction, thus demonstrating the potential of the described system to be modified and adapted for preclinical research.