Evidence map›Paper›PMID 36029346›Full record

ArticleEuropean radiology2023

Reducing cardiac implantable electronic device-induced artefacts in cardiac magnetic resonance imaging.

Aino-Maija Vuorinen, Lauri Lehmonen, Jarkko Karvonen, Miia Holmström, Sari Kivistö, Touko Kaasalainen

Open access · hybridAbstract read
In one paragraph

Article in European radiology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
9citing papers in PubMed, 1 pooled it
2.3field-weighted citation impact, top 12% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

9 citing papers in PubMed, 1 synthesis or guideline pooled it, 17 citations in OpenAlex.

  1. SCMR expert consensus statement for cardiovascular magnetic resonance of patients with a cardiac implantable electronic device.Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance · 2024
    Guideline
  2. Review
  3. Article
  4. Review
  5. Review
  6. Joint suppression of cardiac bSSFP cine banding and flow artifacts using twofold phase-cycling and a dual-encoder neural network.Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance · 2024
    Article
  7. Article
  8. Article
  9. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 1 institution in 1 country.

Aino-Maija VuorinenRadiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, HUS, 00029, Helsinki, Finland. aino-maija.vuorinen@hus.fi.ORCID http://orcid.org/0000-0003-4686-3269
Lauri LehmonenRadiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, HUS, 00029, Helsinki, Finland.
Jarkko KarvonenHeart and Lung Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, HUS, 00029, Helsinki, Finland.
Miia HolmströmRadiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, HUS, 00029, Helsinki, Finland.
Sari KivistöRadiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, HUS, 00029, Helsinki, Finland.
Touko KaasalainenRadiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, HUS, 00029, Helsinki, Finland.
University of Helsinki · FI

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesCardiac implantable electronic device (CIED)-induced metal artefacts possibly significantly diminish the diagnostic value of magnetic resonance imaging (MRI), particularly cardiac MR (CMR). Right-sided generator implantation, wideband late-gadolinium enhancement (LGE) technique and raising the ipsilateral arm to the generator during CMR scanning may reduce the CIED-induced image artefacts. We assessed the impact of generator location and the arm-raised imaging position on the CIED-induced artefacts in CMR.

methodsWe included all clinically indicated CMRs performed on patients with normal cardiac anatomy and a permanent CIED with endocardial pacing leads between November 2011 and October 2019 in our institution (n = 171). We analysed cine and LGE sequences using the American Heart Association 17-segment model for the presence of artefacts.

resultsRight-sided generator implantation and arm-raised imaging associated with a significantly increased number of artefact-free segments. In patients with a right-sided pacemaker, the median percentage of artefact-free segments in short-axis balanced steady-state free precession LGE was 93.8% (IQR 9.4%, n = 53) compared with 78.1% (IQR 20.3%, n = 58) for left-sided pacemaker (p < 0.001). In patients with a left-sided implantable cardioverter-defibrillator, the median percentage of artefact-free segments reached 87.5% (IQR 6.3%, n = 9) using arm-raised imaging, which fell to 62.5% (IQR 34.4%, n = 9) using arm-down imaging in spoiled gradient echo short-axis cine (p = 0.02).

conclusionsArm-raised imaging represents a straightforward method to reduce CMR artefacts in patients with left-sided generators and can be used alongside other image quality improvement methods. Right-sided generator implantation could be considered in CIED patients requiring subsequent CMR imaging to ensure sufficient image quality. KEY POINTS: • Cardiac implantable electronic device (CIED)-induced metal artefacts may significantly diminish the diagnostic value of an MRI, particularly in cardiac MRIs. • Raising the ipsilateral arm relative to the CIED generator is a cost-free, straightforward method to significantly reduce CIED-induced artefacts on cardiac MRIs in patients with a left-sided generator. • Right-sided generator implantation reduces artefacts compared with left-sided implantation and could be considered in CIED patients requiring subsequent cardiac MRIs to ensure adequate image quality in the future.

Indexed as

Defibrillators, ImplantableMagnetic Resonance Imaging, CineContrast MediaGadoliniumHumansMagnetic Resonance ImagingContrast MediaGadoliniumArtefactsDefibrillators, ImplantableElectrodes, ImplantedMagnetic resonance imagingPacemaker, Artificial

Identifiers

PMID36029346
PMCPMC9889467
OpenAlexW4293354865

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.