Evidence map›Paper›PMID 36000957›Full record

ArticleJournal of the Chinese Medical Association : JCMA2022

Optimal depth of electromyographic endotracheal tube: A novel approach using video laryngoscopy.

Yu-Chun Lin, Ming-Chih Lin, Wen-Chun Lin, Ching-Hui Shen

Open access · hybridAbstract read
In one paragraph

Article in Journal of the Chinese Medical Association : JCMA, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.3field-weighted citation impact, top 42% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 2 citations in OpenAlex.

  1. Comparison of EMG responses obtained from endotracheal tube and thyroid cartilage surface electrodes in thyroidectomies performed with intraoperative nerve monitoring.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
    Article
  2. Observational
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 3 institutions in 1 country.

Yu-Chun LinDepartment of Anesthesiology, Taichung Veterans General Hospital, Taichung, Taiwan, ROC.
Ming-Chih LinDepartment of Pediatrics, Taichung Veterans General Hospital, Taichung, Taiwan, ROC.
Wen-Chun LinDepartment of Otolaryngology-Head and Neck Surgery, Taichung Veterans General Hospital, Taichung, Taiwan, ROC.
Ching-Hui ShenDepartment of Anesthesiology, Taichung Veterans General Hospital, Taichung, Taiwan, ROC.
National Chung Hsing University · TWProvidence University · TWTaichung Veterans General Hospital · TW

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundElectromyographic (EMG) endotracheal tubes with surface electrodes are used during neck surgery to prevent recurrent laryngeal nerve (RLN) injury. Proper positioning of the EMG tube is of paramount importance. In this study, we aimed to compare the use of video laryngoscopy with other methods for achieving the optimal depth of the EMG tube.

methodsWe retrospectively enrolled 489 adult patients (with 675 nerves at risk [NAR]) undergoing surgery using the EMG endotracheal tube. Patients were categorized into three groups with: rigid laryngoscope (n = 140, NAR = 187), conventional laryngoscope (n = 262, NAR = 370), and video laryngoscope (n = 87, NAR = 118). A formula for predicting optimal depths of the EMG tube was obtained from data of the standard group with rigid laryngoscope. Depths of the EMG endotracheal tube were measured and postoperative RLN injuries were analyzed.

resultsBased on linear regression, the formula was derived for predicting the optimal depth of EMG endotracheal tube (cm) = 11.028 + 0.635 * gender (female = 0; male = 1) + 0.069 * height (cm). Compared to conventional laryngoscope, intubation of EMG tube with video laryngoscope resulted in less discrepancy between its actual value and optimal value, and the tube depth was more correct (OR = 2.888, 95% CI = 1.753-4.757, p < 0.001). All five postoperative permanent RLN injuries were found in the group with conventional laryngoscope.

conclusionEMG endotracheal tube insertion with video laryngoscopy is superior to conventional laryngoscopy, as well as an alternative to rigid laryngoscopy. The video laryngoscopy is a novel approach to get optimal depth of EMG endotracheal tube during neck surgery.

Indexed as

LaryngoscopesAdultElectrodesFemaleHumansIntubation, IntratrachealLaryngoscopyMaleRetrospective Studies

Identifiers

PMID36000957
PMCPMC12755330
OpenAlexW4292953695

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.