Relapsing low-flow alarms due to suboptimal alignment of the left ventricular assist device inflow cannula.

Casper F Zijderhand, Wiebe G Knol, Ricardo P J Budde, Cornelis W van der Heiden, Kevin M Veen, Jelena Sjatskig, Olivier C Manintveld, Alina A Constantinescu, Ozcan Birim, Jos A Bekkers and 2 more

Abstract read
In one paragraph

Article in European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

12 authors.

Casper F ZijderhandDepartment of Cardiothoracic Surgery, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Wiebe G KnolDepartment of Cardiothoracic Surgery, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Ricardo P J BuddeDepartment of Radiology and Nuclear Medicine, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Cornelis W van der HeidenDepartment of Cardiology, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Kevin M VeenDepartment of Cardiothoracic Surgery, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Jelena SjatskigDepartment of Cardiothoracic Surgery, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Olivier C ManintveldDepartment of Cardiology, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Alina A ConstantinescuDepartment of Cardiology, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Ozcan BirimDepartment of Cardiothoracic Surgery, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.
Jos A BekkersDepartment of Cardiothoracic Surgery, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.ORCID 0000-0002-0221-9239
Ad J J C BogersDepartment of Cardiothoracic Surgery, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.ORCID 0000-0001-6200-1998
Kadir CaliskanDepartment of Cardiology, Thoraxcenter, Erasmus MC, University Medical Center Rotterdam, Rotterdam, Netherlands.ORCID 0000-0002-3293-9261

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesThis retrospective study investigated the correlation between the angular position of the left ventricular assist device (LVAD) inflow cannula and relapsing low-flow alarms.

methodsMedical charts were reviewed of all patients with HeartMate 3 LVAD support for relapsing low-flow alarms. A standardized protocol was created to measure the angular position with a contrast-enhanced computed tomography scan. Statistics were done using a gamma frailty model with a constant rate function.

resultsFor this analysis, 48 LVAD-supported patients were included. The majority of the patients were male (79%) with a median age of 57 years and a median follow-up of 30 months (interquartile range: 19-41). Low-flow alarm(s) were experienced in 30 (63%) patients. Angulation towards the septal-lateral plane showed a significant increase in low-flow alarms over time with a constant rate function of 0.031 increase in low-flow alarms per month of follow-up per increasing degree of angulation (P = 0.048). When dividing this group using an optimal cut-off point, a significant increase in low-flow alarms was observed when the septal-lateral angulation was 28° or more (P = 0.001). Anterior-posterior and maximal inflow cannula angulation did not show a significant difference.

conclusionsThis study showed an increasing number of low-flow alarms when the degrees of LVAD inflow cannula expand towards the septal-lateral plane. This emphasizes the importance of the LVAD inflow cannula angular position to prevent relapsing low-flow alarms with the risk of diminished quality of life and morbidity.

Indexed as

Heart-Assist DevicesHeart FailureCannulaFemaleHeart VentriclesHumansMaleMiddle AgedQuality of LifeRetrospective StudiesTomography, X-Ray ComputedContrast-enhanced computed tomographyHeart failureInflow cannulaLeft ventricular assist deviceLow-flow alarms

Identifiers

PMID35993906
PMCPMC9789739

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.