A fast GPU-accelerated Monte Carlo engine for calculation of MLC-collimated electron fields.

Eric E Brost, H Wan Chan Tseung, John A Antolak

Abstract read
In one paragraph

Article in Medical physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Eric E BrostDepartment of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA.
H Wan Chan TseungDepartment of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA.
John A AntolakDepartment of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA.

Funding

Mayo Clinic School of Graduate Medical Education
6 · The paper itself

Abstract

backgroundAlthough intensity-modulated radiation therapy and volumetric arc therapy have revolutionized photon external beam therapies, the technological advances associated with electron beam therapy have fallen behind. Modern linear accelerators contain technologies that would allow for more advanced forms of electron treatments, such as beam collimation, using the conventional photon multi-leaf collimator (MLC); however, no commercial solutions exist that calculate dose from such beam delivery modes. Additionally, for clinical adoption to occur, dose calculation times would need to be on par with that of modern dose calculation algorithms. PURPOSE: This work developed a graphics processing unit (GPU)-accelerated Monte Carlo (MC) engine incorporating the Varian TrueBeam linac head geometry for a rapid calculation of electron beams collimated using the conventional photon MLC.

methodsA compute unified device architecture framework was created for the following: (1) transport of electrons and photons through the linac head geometry, considering multiple scattering, Bremsstrahlung, Møller, Compton, and pair production interactions; (2) electron and photon propagation through the CT geometry, considering all interactions plus the photoelectric effect; and (3) secondary particle cascades through the linac head and within the CT geometry. The linac head collimating geometry was modeled according to the specifications provided by the vendor, who also provided phase-space files. The MC was benchmarked against EGSnrc/DOSXYZnrc/GEANT by simulating individual interactions with simple geometries, pencil, and square beam dose calculations in various phantoms. MC-calculated dose distributions for MLC and jaw-collimated electron fields were compared to measurements in a water phantom and with radiochromic film.

resultsPencil and square beam dose distributions are in good agreement with DOSXYZnrc. Angular and spatial distributions for multiple scattering and secondary particle production in thin slab geometries are in good agreement with EGSnrc and GEANT. Dose profiles for MLC and jaw-collimated 6-20-MeV electron beams showed an average absolute difference of 1.1 and 1.9 mm for the FWHM and 80%-20% penumbra from measured profiles. Percent depth doses showed differences of <5% for as compared to measurement. The computation time on an NVIDIA Tesla V100 card was 2.5 min to achieve a dose uncertainty of <1%, which is ∼300 times faster than published results in a similar geometry using a single-CPU core.

conclusionsThe GPU-based MC can quickly calculate dose for electron fields collimated using the conventional photon MLC. The fast calculation times will allow for a rapid calculation of electron fields for mixed photon and electron particle therapy.

Indexed as

ElectronsRadiotherapy, Intensity-ModulatedAlgorithmsMonte Carlo MethodParticle AcceleratorsPhantoms, ImagingPhotonsRadiotherapy DosageRadiotherapy Planning, Computer-AssistedCUDAelectron-beam therapyGPUMonte Carlo

Identifiers

PMID35986907
PMCPMC10087940

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.