Monte Carlo modeling of the Elekta Versa HD and patient dose calculation with EGSnrc/BEAMnrc.
Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Dosimetric investigation of small fields in radiotherapy measurements using Monte Carlo simulations, CC04 ionization chamber, and razor diode.Physical and engineering sciences in medicine · 2025Article
- Monte Carlo modeling of the Elekta Versa HD and patient dose calculation with EGSnrc/BEAMnrc.Journal of applied clinical medical physics · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
introductionNumerous studies have proven the Monte Carlo method to be an accurate means of dose calculation. Although there are several commercial Monte Carlo treatment planning systems (TPSs), some clinics may not have access to these resources. We present a method for routine, independent patient dose calculations from treatment plans generated in a commercial TPS with our own Monte Carlo model using free, open-source software. MATERIALS AND
methodsA model of the Elekta Versa HD linear accelerator was developed using the EGSnrc codes. A MATLAB script was created to take clinical patient plans and convert the DICOM RTP files into a format usable by EGSnrc. Ten patients' treatment plans were exported from the Monaco TPS to be recalculated using EGSnrc. Treatment simulations were done in BEAMnrc, and doses were calculated using Source 21 in DOSXYZnrc. Results were compared to patient plans calculated in the Monaco TPS and evaluated in Verisoft with a gamma criterion of 3%/2 mm.
resultsOur Monte Carlo model was validated within 1%/1-mm accuracy of measured percent depth doses and profiles. Gamma passing rates ranged from 82.1% to 99.8%, with 7 out of 10 plans having a gamma pass rate over 95%. Lung and prostate patients showed the best agreement with doses calculated in Monaco. All statistical uncertainties in DOSXYZnrc were less than 3.0%.
conclusionA Monte Carlo model for routine patient dose calculation was successfully developed and tested. This model allows users to directly recalculate DICOM RP files containing patients' plans that have been exported from a commercial TPS.