ArticleNeuroImage2022
Cross-Axis Dynamic Field Compensation of Optically Pumped Magnetometer Arrays for MEG.
Article in NeuroImage, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
12 citing papers in PubMed, 48 citations in OpenAlex.
- Human cardiac measurements with diamond magnetometers.Science advances · 2026Article
- Environment-Driven Synthetic Baseline Analysis and Optimization in Joint Measurement OPM-MEG Arrays.Bioengineering (Basel, Switzerland) · 2026Article
- Optimization and Analysis of Tangential Component Orientations in OPM-MEG Sensor Array.Bioengineering (Basel, Switzerland) · 2025Article
- Combining video telemetry and wearable MEG for naturalistic imaging.Imaging neuroscience (Cambridge, Mass.) · 2025Article
- Inferring laminar origins of MEG signals with optically pumped magnetometers (OPMs): A simulation study.Imaging neuroscience (Cambridge, Mass.) · 2025Article
- Gradient phase and amplitude errors in atomic magnetic gradiometers for biomagnetic imaging systems.iScience · 2024Article
- Adaptive multipole models of optically pumped magnetometer data.Human brain mapping · 2024Article
- Simultaneous whole-head electrophysiological recordings using EEG and OPM-MEG.Imaging neuroscience (Cambridge, Mass.) · 2024Article
- Article
- An integrated full-head OPM-MEG system based on 128 zero-field sensors.Frontiers in neuroscience · 2023Article
- Optimizing NV magnetometry for Magnetoneurography and Magnetomyography applications.Frontiers in neuroscience · 2022Article
- Channel gain calibrated signal space projection using variational Bayesian estimation for optically pumped magnetometers.Imaging neuroscience (Cambridge, Mass.)Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors at 2 institutions in 1 country.
Funding
Abstract
We present dynamic field compensation (DFC), whereby three-axis field measurements from reference magnetometers are used to dynamically maintain null at the alkali vapor cells of an array of primary sensors that are proximal to a subject's scalp. Precision measurement of the magnetoencephalogram (MEG) by zero-field optically pumped magnetometer (OPM) sensors requires that sensor response is linear and sensor gain is constant over time. OPMs can be operated in open-loop mode, where the measured field is proportional to the output at the demodulated photodiode output, or in closed-loop, where on-board coils are dynamically driven to maintain the internal cell at zero field in the measurement direction. While OPMs can be operated in closed-loop mode along all three axes, this can increase sensor noise and poses engineering challenges. Uncompensated fluctuations in the ambient field along any statically nulled axes perturb the measured field by tipping the measurement axis and altering effective sensor gain - a phenomenon recently referred to as cross-axis projection error (CAPE). These errors are particularly problematic when OPMs are allowed to move in the remnant background field. Sensor gain-errors, if not mitigated, preclude precision measurements with OPMs operating in the presence of ambient field fluctuations within a typical MEG laboratory. In this manuscript, we present the cross-axis dynamic field compensation (DFC) method for maintaining zero field dynamically on all three axes of each sensor in an array of OPMs. Together, DFC and closed-loop operation strongly attenuate errors introduced by CAPE. This method was implemented by using three orthogonal reference sensors together with OPM electronics that permit driving each sensor's transverse field coils dynamically to maintain null field across its OPM measurement cell. These reference sensors can also be used for synthesizing 1