Evidence map›Paper›PMID 35956198›Full record

ArticleJournal of clinical medicine2022

Assessment of the New Acceleromyograph TOF 3D Compared with the Established TOF Watch SX: Bland-Altman Analysis of the Precision and Limits of Agreement between Both Devices-A Randomized Clinical Comparison.

Stefan Soltesz, Jan Thomas, Michael Anapolski, Guenter Karl Noé

Open access · goldAbstract read
In one paragraph

Article in Journal of clinical medicine, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
1.0field-weighted citation impact, top 23% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 5 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 1 institution in 1 country.

Stefan SolteszDepartment of Anesthesia and Intensive Care Medicine, Rheinlandklinikum Dormagen, 41540 Dormagen, Germany.ORCID 0000-0001-6460-6878
Jan ThomasDepartment of Anesthesia and Intensive Care Medicine, Rheinlandklinikum Dormagen, 41540 Dormagen, Germany.
Michael AnapolskiDepartment Ob/Gyn, Rheinlandklinikum Dormagen, University of Witten-Herdecke, 41540 Dormagen, Germany.
Guenter Karl NoéDepartment Ob/Gyn, Rheinlandklinikum Dormagen, University of Witten-Herdecke, 41540 Dormagen, Germany.
Kreiskrankenhaus Dormagen · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The new acceleromyograph TOF 3D was compared with the established TOF Watch SX in patients undergoing elective laparoscopic gynecological surgery. Neuromuscular transmission was assessed by simultaneous recording with both devices. Measurements were performed simultaneously at the left and the right M. adductor pollicis (Group A, 25 patients), or the M. corrugator supercilii (Group CS, 25 patients). The repeatability, time course, and limits of agreement (Bland-Altman) were compared. The primary endpoint was the 90% train-of-four recovery time (TOFR 0.9). Other endpoints included onset time of block, maximum T1 depression, time to 25% T1 recovery, the recovery time course of T1 response, and TOF ratio, respectively. In group CS, the repeatability coefficient of the TOF 3D was lower (4.66 (1.6)) than of the TOF Watch SX (6.02 (1.9);

Indexed as

acceleromyographyneuromuscular blockneuromuscular monitoring

Identifiers

PMID35956198
PMCPMC9369732
OpenAlexW4290573470

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked