Comparison of the visual field test of Glaufield Lite with Humphrey Field Analyser.
Article in International ophthalmology, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo compare visual field test results of Glaufield Lite AP901 CTS 133 (Appasamy Associates, Mannadipet Commune, Thirubhuvanai, Puducherry, India, hereafter Glaufield Lite) with Humphrey Field Analyser (HFA, Carl Zeiss Meditec, Dublin, California, USA, hereafter HFA).
methodsA pilot study at a tertiary eye centre involving 23 normal and 24 glaucoma patients who underwent two consecutive visual field tests on (i) HFA 24-2 SITA Fast and (ii) Glaufield Lite Quick Central program.
resultsThe mean testing time on HFA was significantly shorter than Glaufield Lite (normals: HFA: 2.75 ± 0.49 min, Glaufield Lite: 6.85 ± 0.86 min, p < 0.001; glaucoma patients: HFA: 3.45 ± 1.08 min, Glaufield Lite: 6.95 ± 0.54 min, p < 0.001). Reliability criteria were similar, but false-positivity was lower with Glaufield Lite. Bland-Altman analysis showed poor agreement for mean deviation (MD), [~ 2.69 units less for HFA], and acceptable agreement for pattern standard deviation (PSD) [~ 0.426 units more for HFA] between the two devices.
conclusionBoth perimetric techniques showed reliable test results though test duration was longer with Glaufield Lite perimetry. The MD showed poor agreement, likely due to different scales and principles used for perimetry. The PSD showed acceptable agreement, making it valid for use in glaucoma, though a direct comparison of fields from the two devices is not possible. We recommend using the same perimetry device for follow-up evaluation.