Evidence map›Paper›PMID 35869775›Full record

ArticleNan fang yi ke da xue xue bao = Journal of Southern Medical University2022

[Impact of multi-leaf collimator positioning accuracy on quality control of volumetric modulation arc therapy plan for cervical cancer treated with Elekta linear accelerator].

J Liang, F Yu, J Zhu, T Song

Open access · greenAbstract read
In one paragraph

Article in Nan fang yi ke da xue xue bao = Journal of Southern Medical University, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact, top 90% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed, 0 citations in OpenAlex.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 1 institution in 1 country.

J LiangSchool of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China.
F YuState Key Laboratory of Oncology in South China//Department of Radiotherapy, Sun Yat-sen University Cancer Center, Guangzhou 510060, China.
J ZhuState Key Laboratory of Oncology in South China//Department of Radiotherapy, Sun Yat-sen University Cancer Center, Guangzhou 510060, China.
T SongSchool of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China.
Southern Medical University · CN

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveTo investigate the influence of positioning accuracy of the multi-leaf collimators (MLC) on the passing rate of the plan dose verification for volumetric modulation arc therapy (VMAT) of cervical cancer using an Elekta linear accelerator.

methodsThe dose distributions were measured using Sun Nuclear's Mapcheck and Arccheck semiconductors matrix before and after MLC calibration in30 cervical cancer patients undergoing VMAT. Dosimetric comparisons were performed with 2D and 3D gamma passing rates of 3%, 3 mm and 2%, and 2 mm. The 3D gamma distribution was reconstructed with respect to the patient's anatomy using 3DVH software to evaluate the possible influence of MLC positioning accuracy.

resultsBefore and after MLC calibration, the gamma passing rates of Mapcheck were (88.80±1.81)% and (99.25 ± 0.53)% under 3% and 3 mm standard, respectively, with an average increase of 10.45%. The corresponding gamma passing rates of Arccheck were (87.61±1.98)% and (98.13±0.99)%, respectively, with an average increase of 10.52%. The gamma passing rates of 3DVH were (89.87±2.28)% and (98.3±1.15)%, respectively, with an average increase of 8.43%.

conclusionThe MLC positioning accuracy is one of the main factors influencing dosimetric accuracy of VMAT for cervical cancer. The application of Autocal software facilitates MLC calibration and improves the accuracy and safety of VMAT delivery for cervical cancer.

Indexed as

Radiotherapy, Intensity-ModulatedUterine Cervical NeoplasmsFemaleHumansParticle AcceleratorsQuality ControlRadiotherapy DosageRadiotherapy Planning, Computer-AssistedAutocalcervical cancermulti-leaf collimatorplan verificationvolumetric modulation arc therapy

Identifiers

PMID35869775
PMCPMC9308871
OpenAlexW4286818707

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.