ArticleJournal of applied clinical medical physics2022
Analysis of a surface imaging system using a six degree-of-freedom couch.
Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed.
- Frameless linac-based radiosurgery for benign intracranial tumors treated with HyperArc: analysis of tumor control and toxicity.Journal of neuro-oncology · 2025Article
- New findings on clinical experience on surface-guided radiotherapy for frameless non-coplanar stereotactic radiosurgery treatments.Journal of applied clinical medical physics · 2024Article
- Article
- Characterization of the IDENTIFYJournal of applied clinical medical physics · 2024Article
- Evaluation and correlation of patient movement during online adaptive radiotherapy with CBCT and a surface imaging system.Journal of applied clinical medical physics · 2023Article
- Commissioning and clinical evaluation of the IDENTIFYJournal of applied clinical medical physics · 2023Article
- A roadmap for implementation of kV-CBCT online adaptive radiation therapy and initial first year experiences.Journal of applied clinical medical physics · 2023Article
- Analysis of a surface imaging system using a six degree-of-freedom couch.Journal of applied clinical medical physics · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo validate surface imaging (SI)-reported offsets using a six degree-of-freedom couch and an anthropomorphic phantom for commissioning and routine quality assurance of an SI system used for stereotactic radiosurgery (SRS).
methodsAn anthropomorphic phantom with a radiopaque ball bearing (BB) placed either anterior, midline, or posterior, was tracked with SI with a typical SRS region of interest. Couch motion in all six degrees of freedom was programmed and delivered on a linac. SI system logs were synchronized with linac trajectory logs. Ten random couch positions were selected at couch 0°, 45°, 90°, 270°, 315° with megavolt (MV) images taken to account for couch walkout. The SI residual error (ε), the difference between SI reported offset and MV or trajectory log position, was calculated. Residual errors were measured with and without one SI pod blocked.
resultsThe median [range] of magnitude of translational ε was 0.13 [0.07, 0.21], 0.16 [0.11, 0.26], 0.61 [0.50, 0.68], 0.49 [0.42, 0.55], 0.55 [0.38, 0.72] mm for couch rotations of 0°, 45°, 90°, 270°, 315°, respectively, for the midline BB and no pod blocked. The range of all translational ε from all couch angles (with and without pod block) at different BB positions is [0.05, 0.96] mm. The absolute range of difference when changing BB position when no pod is blocked in median translational ε is [0.01, 0.40] mm with the maximum at BB posterior. The absolute range of difference when not changing BB positions with and without pod block in median translational ε is [0.01, 0.37] mm with the maximum at BB posterior and couch 315°.
conclusionSI system and linac trajectory log analysis can be used to assess SI system performance with automated couch motion to validate SI accuracy.