Evidence map›Paper›PMID 35819580›Full record

ArticleCardiovascular engineering and technology2022

Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold.

Ben Hoddy, Naveed Ahmed, Kadem Al-Lamee, Nial Bullett, Nick Curzen, Neil W Bressloff

Open access · hybridAbstract read
In one paragraph

Article in Cardiovascular engineering and technology, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.2field-weighted citation impact, top 47% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 1 citations in OpenAlex.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 2 institutions in 1 country.

Ben HoddyComputational Engineering and Design Research Group, University of Southampton, Southampton, UK. bh1g12@soton.ac.uk.ORCID 0000-0002-3918-9977
Naveed AhmedArterius Ltd, Leeds, UK.
Kadem Al-LameeArterius Ltd, Leeds, UK.
Nial BullettArterius Ltd, Leeds, UK.
Nick CurzenCoronary Research Group, Southampton University Hospitals NHS Trust, Southampton, UK.
Neil W BressloffComputational Engineering and Design Research Group, University of Southampton, Southampton, UK.
Arterius (United Kingdom) · GBUniversity of Southampton · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe ArterioSorb[Formula: see text] bioresorbable scaffold (BRS) developed by Arterius Ltd is about to enter first in man clinical trials. Previous generations of BRS have been vulnerable to brittle fracture, when expanded via balloon inflation in-vivo, which can be extremely detrimental to patient outcome. Therefore, this study explores the effect of variable ring length and strut width (as facilitated by the ArterioSorb[Formula: see text] design) on fracture resistance via analysis of the distribution of equivalent plastic strain in the scaffold struts post expansion. Scaffold performance is also assessed with respect to side branch access, radial strength, final deployed diameter and percentage recoil.

methodsFinite element analysis was conducted of the crimping, expansion and radial crushing of five scaffold designs comprising different variations in ring length and strut width. The Abaqus/Explicit (DS SIMULIA) solution method was used for all simulations. Direct comparison between in-silico predictions and in-vitro measurements of the performance of the open cell variant of the ArterioSorb[Formula: see text] were made. Paths across the width of the crown apex and around the scaffold rings were defined along which the plastic strain distribution was analysed.

resultsThe in-silico results demonstrated good predictions of final shape for the baseline scaffold design. Percentage recoil and radial strength were predicted to be, respectively, 2.8 and 1.7 times higher than the experimentally measured values, predominantly due to the limitations of the anisotropic elasto-plastic material property model used for the scaffold. Average maximum values of equivalent plastic strain were up to 2.4 times higher in the wide strut designs relative to the narrow strut scaffolds. As well as the concomitant risk of strut fracture, the wide strut designs also exhibited twisting and splaying behaviour at the crowns located on the scaffold end rings. Not only are these phenomena detrimental to the radial strength and risk of strut fracture but they also increase the likelihood of damage to the vessel wall. However, the baseline scaffold design was observed to tolerate significant over expansion without inducing excessive plastic strains, a result which is particularly encouraging, due to post-dilatation being commonplace in clinical practice.

conclusionTherefore, the narrow strut designs investigated herein, are likely to offer optimal performance and potentially better patient outcomes. Further work should address the material modelling of next generation polymeric BRS to more accurately capture their mechanical behaviour. Observation of the in-vitro testing indicates that the ArterioSorb[Formula: see text] BRS can tolerate greater levels of over expansion than anticipated.

Indexed as

Absorbable ImplantsPercutaneous Coronary InterventionHumansPlasticsProsthesis DesignTomography, Optical CoherencePlasticsBioresorbableFEAPEEQPLLAScaffold

Identifiers

PMID35819580
PMCPMC9750924
OpenAlexW4285044453

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.