ArticleJournal of applied clinical medical physics2022

Automated development of the contrast-detail curve based on statistical low-contrast detectability in CT images.

Choirul Anam, Ariij Naufal, Toshioh Fujibuchi, Kosuke Matsubara, Geoff Dougherty

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Choirul AnamDepartment of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Central Java, Indonesia.
Ariij NaufalDepartment of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Central Java, Indonesia.
Toshioh FujibuchiDepartment of Health Sciences, Faculty of Medical Sciences, Kyushu University, Fukuoka, Japan.
Kosuke MatsubaraDepartment of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Japan.
Geoff DoughertyDepartment of Applied Physics and Medical Imaging, California State University Channel Islands, Camarillo, California, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeWe have developed a software to automatically find the contrast-detail (C-D) curve based on the statistical low-contrast detectability (LCD) in images of computed tomography (CT) phantoms at multiple cell sizes and to generate minimum detectable contrast (MDC) characteristics.

methodsA simple graphical user interface was developed to set the initial parameters needed to create multiple grid region of interest of various cell sizes with a 2-pixel increment. For each cell in the grid, the average CT number was calculated to obtain the standard deviation (SD). Detectability was then calculated by multiplying the SD of the mean CT numbers by 3.29. This process was automatically repeated as many times as the cell size was set at initialization. Based on the obtained LCD, the C-D curve was obtained and the target size at an MDC of 0.6% (i.e., 6-HU difference) was determined. We subsequently investigated the consistency of the target sizes for a 0.6% MDC at four locations within the homogeneous image. We applied the software to images with six noise levels, images of two modules of the American College of Radiology CT phantom, images of four different phantoms, and images of four different CT scanners. We compared the target sizes at a 0.6% MDC based on the statistical LCD and the results from a human observer.

resultsThe developed system was able to measure C-D curves from different phantoms and scanners. We found that the C-D curves follow a power-law fit. We found that higher noise levels resulted in a higher MDC for a target of the same size. The low-contrast module image had a slightly higher MDC than the distance module image. The minimum size of an object detected by visual observation was slightly larger than the size using statistical LCD.

conclusionsThe statistical LCD measurement method can generate a C-D curve automatically, quickly, and objectively.

Indexed as

SoftwareTomography, X-Ray ComputedAlgorithmsHumansPhantoms, ImagingRadiation DosageTomography Scanners, X-Ray Computedimage qualitylow-contrast detectabilityminimum detectable contrast

Identifiers

PMID35808971
PMCPMC9512356

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.