ArticlePediatric pulmonology2022
Breath-dependent pressure fluctuations in various constant- and variable-flow neonatal CPAP devices.
Article in Pediatric pulmonology, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed, 8 citations in OpenAlex.
- Neonatal bench study of a constant-flow device.Pediatric research · 2026Article
- Imposed work of breathing of 16 neonatal CPAP-devices using different mechanisms of CPAP generation.Pediatric research · 2026Article
- What CPAP to use in the delivery room? Bench comparison of two methods to provide continuous positive airways pressure in neonates.BMJ paediatrics open · 2024Article
- Function of neonatal CPAP systems and the importance of total flow.Pediatric pulmonology · 2023Article
- Breath-dependent pressure fluctuations in various constant- and variable-flow neonatal CPAP devices.Pediatric pulmonology · 2022Article
Corrections and comments
- Commented on by
Authors and funding
6 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveIn continuous positive airway pressure (CPAP) devices, pressure can be generated by two different mechanisms: either via an expiratory valve or by one or more jets. Valved CPAP devices are referred to as constant-flow devices, and jet devices are called variable-flow devices. Constant-flow CPAP devices are said to reduce the imposed work of breathing due to lower breath-dependent pressure fluctuations. The present study investigates the performance of various constant- and variable-flow CPAP devices in relation to breath-dependent pressure fluctuations.
designExperimental study comparing the pressure fluctuations incurred by seven neonatal CPAP devices attached to an active neonatal lung model. METHODOLOGY: Spontaneous breathing was simulated using a tidal volume of 6 ml at pressure levels of 5, 7, and 9 mbar. The main outcomes were respiratory pressure fluctuations, tidal volume, and end-expiratory pressure.
resultsAll CPAP devices tested showed respiratory pressure fluctuations, varying from 0.631 to 3.466 mbar. The generated tidal volume correlated significantly with the pressure fluctuations (r = -0.947; p = 0.001) and varied between 5.550 and 6.316 ml. CPAP devices with jets showed no advantage over CPAP devices with expiratory valves. End-expiratory pressure in the nose deviated from the set pressure between -1.305 and 0.644 mbar and varied depending on whether the pressure was measured in the device or in the tube extending to the nose.
conclusionDuring standard spontaneous breathing, breath-dependent pressure fluctuations in constant- and variable-flow devices are comparable. Pressure measurements taken in the tubing system can lead to a considerable deviation of the applied pressure.