ArticleMedical & biological engineering & computing2022
Verifying a C-arm-based roentgen stereophotogrammetric analysis protocol for assessing tibial implant movement in total knee arthroplasty.
Article in Medical & biological engineering & computing, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Roentgen stereophotogrammetric analysis (RSA) is the "gold standard" technique for measuring sub-millimetric relative motion between implant and bone to quantify post-operative implant migration over time. The vast majority of RSA studies addressing implant motion in knee replacements, however, have been conducted using expensive biplanar radiography systems and commercial software that are not readily available at many institutions. In this study, we evaluated the feasibility of performing RSA using ordinary, readily available C-arm fluoroscopes and open-source software to assess tibial component migration.We developed an assessment protocol using a Siemens Arcadis Orbic C-arm and the open-source XROMM software and evaluated its accuracy and precision through a series of phantom-based verification tests. The results were highly promising: accuracies were in the range of - 39 to 11 μm for translations and - 0.025 to 0.029° for rotations, while system precisions ranged between 16 to 27 μm for translations and 0.041 to 0.059° for rotations. This performance is comparable to specialized RSA systems reported in the literature. The proposed RSA protocol is therefore capable of accurately measuring the relative motion of knee replacement implants in phantom scenarios, which justifies further the development of the protocol towards use in prospective clinical assessments of new implant designs and surgical techniques.