Comprehensive evaluation of dosimetric impact against position errors in accelerator-based BNCT under different treatment parameter settings.

Ryo Kakino, Naonori Hu, Kayako Isohashi, Teruhito Aihara, Keiji Nihei, Koji Ono

Abstract read
In one paragraph

Article in Medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed.

  1. Article
  2. Review
  3. Hypo-segmentation ofEJNMMI research · 2025
    Article
  4. Article
  5. Article
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Ryo KakinoKansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Takatsuki-shi, Osaka, Japan.
Naonori HuKansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Takatsuki-shi, Osaka, Japan.
Kayako IsohashiKansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Takatsuki-shi, Osaka, Japan.
Teruhito AiharaKansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Takatsuki-shi, Osaka, Japan.
Keiji NiheiKansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Takatsuki-shi, Osaka, Japan.
Koji OnoKansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Takatsuki-shi, Osaka, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundPatients who undergo accelerator-based (AB) boron neutron capture therapy (BNCT) for head and neck cancer in the sitting position are generally uncomfortably immobilized, and patient motion during this treatment may be greater than that in other radiotherapy techniques. Furthermore, the treatment time of BNCT is relatively long (up to approximately 1 h), which increases the possibility of patient movement during treatment. As most BNCT irradiations are performed in a single fraction, the dosimetric error due to patient motion is of greater consequence and needs to be evaluated and accounted for. Several treatment parameters are required for BNCT dose calculation. PURPOSE: To investigate the dosimetric impacts (DIs) against position errors using a simple cylindrical phantom for an AB-BNCT system under different treatment parameter settings.

methodsThe treatment plans were created in RayStation and the dose calculation was performed using the NeuCure® dose engine. A cylindrical phantom (16 cm diameter × 20 cm height) made of soft tissue was modeled. Dummy tumors in the form of a 3-cm-diameter sphere were arranged at depths of 2.5 and 6.5 cm (denoted by T

resultsThe coefficients of the MRA of the DIs for LR, AP, and SI shifts were -0.08, 2.16, and -0.04 (p-values = 0.084, <0.01, and 0.334) for T

conclusionThe DIs were largely affected by the shift in the AP direction and were influenced by the different treatment parameters.

Indexed as

Boron Neutron Capture TherapyHumansMonte Carlo MethodPhantoms, ImagingRadiometryRadiotherapy Dosageboron neutron capture therapycyclotron-based epithermal neutron sourceMonte Carlo simulationpatient setup error