Evidence map›Paper›PMID 35735560›Full record

ArticleBiosensors2022

ImmunoDisk-A Fully Automated Bead-Based Immunoassay Cartridge with All Reagents Pre-Stored.

Benita Johannsen, Desirée Baumgartner, Lena Karkossa, Nils Paust, Michal Karpíšek, Nagihan Bostanci, Roland Zengerle, Konstantinos Mitsakakis

Open access · goldAbstract read
In one paragraph

Article in Biosensors, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
0.8field-weighted citation impact, top 33% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 11 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 4 institutions in 3 countries.

Benita JohannsenHahn-Schickard, Georges-Koehler-Allee 103, 79110 Freiburg, Germany.
Desirée BaumgartnerLaboratory for MEMS Applications, IMTEK-Department of Microsystems Engineering, University of Freiburg, Georges-Koehler-Allee 103, 79110 Freiburg, Germany.
Lena KarkossaHahn-Schickard, Georges-Koehler-Allee 103, 79110 Freiburg, Germany.
Nils PaustHahn-Schickard, Georges-Koehler-Allee 103, 79110 Freiburg, Germany.
Michal KarpíšekBioVendor-Laboratorní Medicína a.s., Research & Diagnostic Products Division, Karasek 1767/1, Reckovice, 62100 Brno, Czech Republic.ORCID 0000-0001-6503-4250
Nagihan BostanciSection of Oral Health and Periodontology, Division of Oral Diseases, Department of Dental Medicine, Karolinska Institutet, 14104 Huddinge, Sweden.
Roland ZengerleHahn-Schickard, Georges-Koehler-Allee 103, 79110 Freiburg, Germany.
Konstantinos MitsakakisHahn-Schickard, Georges-Koehler-Allee 103, 79110 Freiburg, Germany.ORCID 0000-0002-3287-9123
University of Freiburg · DEHahn-Schickard-Gesellschaft für angewandte Forschung · DEKarolinska Institutet · SEMasaryk University · CZ

Funding

European Commission 633780Eurostars, managed by the Federal Ministry of Education & Research (BMBF) E!113595
6 · The paper itself

Abstract

In this paper, we present the ImmunoDisk, a fully automated sample-to-answer centrifugal microfluidic cartridge, integrating a heterogeneous, wash-free, magnetic- and fluorescent bead-based immunoassay (bound-free phase detection immunoassay/BFPD-IA). The BFPD-IA allows the implementation of a simple fluidic structure, where the assay incubation, bead separation and detection are performed in the same chamber. The system was characterized using a C-reactive protein (CRP) competitive immunoassay. A parametric investigation on air drying of protein-coupled beads for pre-storage at room temperature is presented. The key parameters were buffer composition, drying temperature and duration. A protocol for drying two different types of protein-coupled beads with the same temperature and duration using different drying buffers is presented. The sample-to-answer workflow was demonstrated measuring CRP in 5 µL of human serum, without prior dilution, utilizing only one incubation step, in 20 min turnaround time, in the clinically relevant concentration range of 15-115 mg/L. A reproducibility assessment over three disk batches revealed an average signal coefficient of variation (CV) of 5.8 ± 1.3%. A CRP certified reference material was used for method verification with a concentration CV of 8.6%. Our results encourage future testing of the CRP-ImmunoDisk in clinical studies and its point-of-care implementation in many diagnostic applications.

Indexed as

C-Reactive ProteinMicrofluidicsHumansImmunoassayIndicators and ReagentsReproducibility of ResultsC-Reactive ProteinIndicators and Reagentsbound-free phasecentrifugal microfluidicsimmunoassayinflammationmicro/nanoparticlespoint-of-carereagent storage

Identifiers

PMID35735560
PMCPMC9221266
OpenAlexW4282944234

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked