Instrument tables equipped with local unidirectional airflow units reduce bacterial contamination during orthopedic implant surgery in an operating room with a displacement ventilation system.

Josefin Seth Caous, Karin Svensson Malchau, Max Petzold, Ylva Fridell, Henrik Malchau, Linda Ahlstrom, Peter Grant, Annette Erichsen Andersson

Open access · goldAbstract read
In one paragraph

Article in Infection prevention in practice, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.9field-weighted citation impact, top 28% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 4 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 5 institutions in 3 countries.

Josefin Seth CaousDepartment of Methodology, Textiles and Medical Technology, RISE Research Institutes of Sweden, Gothenburg, Sweden.
Karin Svensson MalchauDepartment of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Sweden.
Max PetzoldSchool of Public Health and Community Medicine, Institute of Medicine, Sahlgrenska Academy, University of Gothenburg, Sweden.
Ylva FridellDepartment of Orthopedics, Sahlgrenska University Hospital, Gothenburg, Sweden.
Henrik MalchauDepartment of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Sweden.
Linda AhlstromDepartment of Orthopedics, Sahlgrenska University Hospital, Gothenburg, Sweden.
Peter GrantDepartment of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Sweden.
Annette Erichsen AnderssonDepartment of Orthopedics, Sahlgrenska University Hospital, Gothenburg, Sweden.
Sahlgrenska University Hospital · SEHarvard University · USLovisenberg Diakonale Sykehus · NORISE Research Institutes of Sweden · SEUniversity of Gothenburg · SE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Airborne bacteria present in the operating room may be a cause of surgical site infection, either contaminating the surgical wound directly, or indirectly via e.g. surgical instruments. The aim of this study was to evaluate if instrument and assistant tables equipped with local unidirectional airflow reduce bacterial contamination of the instrument area to ultra clean levels, during orthopedic implant surgery in an operating room with displacement ventilation. Methods: Local airflow units of instrument and assistant tables were either active or inactive. Colony forming units were sampled intraoperatively from the air above the instruments and from instrument dummies. A minimum of three air samples and two-three samples from instrument dummies were taken during each surgery. Samples were incubated on agar for total aerobic bacterial count. The mean air and instrument contamination during each surgery was calculated and used to analyze the difference in contamination depending on use of local airflow or not. All procedures were performed in the same OR. Results: 188 air and 124 instrument samples were collected during 48 orthopedic implant procedures. Analysis showed that local unidirectional airflow above the surgical instruments significantly reduced the bacterial count in the air above assistant table ( Conclusions: Instrumentation tables equipped with local unidirectional airflow protect the surgical instruments from bacterial contamination during orthopedic implant surgery and may therefore reduce the risk of indirect wound contamination.

Indexed as

Bacterial countColony forming unitInstrument tableOrthopedic surgerySurgical instrumentsUnidirectional airflow