Insensitivity of machine log files to MLC leaf backlash and effect of MLC backlash on clinical dynamic MLC motion: An experimental investigation.

Michael Barnes, Dennis Pomare, Marcus Doebrich, Therese S Standen, Joshua Wolf, Peter Greer, John Simpson

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.

0numbers the graph read from it
0cells of the map it votes in
10citing papers in PubMed
4.0field-weighted citation impact, top 6% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

10 citing papers in PubMed, 15 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Review
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 3 institutions in 2 countries.

Michael BarnesDepartment of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
Dennis PomareDepartment of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
Marcus DoebrichDepartment of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
Therese S StandenDepartment of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
Joshua WolfDepartment of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
Peter GreerDepartment of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
John SimpsonDepartment of Radiation Oncology, Calvary Mater Hospital Newcastle, Newcastle, New South Wales, Australia.
Calvary Mater Newcastle Hospital · AUSouthland Hospital · NZUniversity of Newcastle Australia · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeMulti-leaf-collimator (MLC) leaf position accuracy is important for accurate dynamic radiotherapy treatment plan delivery. Machine log files have become widely utilized for quality assurance (QA) of such dynamic treatments. The primary aim is to test the sensitivity of machine log files in comparison to electronic portal imaging device (EPID)-based measurements to MLC position errors caused by leaf backlash. The secondary aim is to investigate the effect of MLC leaf backlash on MLC leaf motion during clinical dynamic plan delivery.

methodsThe sensitivity of machine log files and two EPID-based measurements were assessed via a controlled experiment, whereby the length of the "T" section of a series of 12 MLC leaf T-nuts in a Varian Millennium MLC for a Trilogy C-series type linac was reduced by sandpapering the top of the "T" to introduce backlash. The built-in machine MLC leaf backlash test as well as measurements for two EPID-based dynamic MLC positional tests along with log files were recorded pre- and post-T-nut modification. All methods were investigated for sensitivity to the T-nut change by assessing the effect on measured MLC leaf positions. A reduced version of the experiment was repeated on a TrueBeam type linac with Millennium MLC.

resultsNo significant differences before and after T-nut modification were detected in any of the log file data. Both EPID methods demonstrated sensitivity to the introduced change at approximately the expected magnitude with a strong dependence observed with gantry angle. EPID-based data showed MLC positional error in agreement with the micrometer measured T-nut length change to 0.07 ± 0.05 mm (1 SD) using the departmental routine QA test. Backlash results were consistent between linac types.

conclusionMachine log files appear insensitive to MLC position errors caused by MLC leaf backlash introduced via the T-nut. The effect of backlash on clinical MLC motions is heavily gantry angle dependent.

Indexed as

Radiotherapy, Intensity-ModulatedElectrical Equipment and SuppliesHumansParticle AcceleratorsPhantoms, ImagingPlant LeavesRadiotherapy Planning, Computer-AssistedEPIDlog filesMLCquality assurance (QA)

Identifiers

PMID35678793
PMCPMC9512360
OpenAlexW4281848098

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.