Split-Window OCT biometry in pseudophakic eyes.
Article in Acta ophthalmologica, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 1 citations in OpenAlex.
- Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer.Eye and vision (London, England) · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors at 2 institutions in 3 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo determine the utility of Split-Window optical coherence tomography OCT (SW-OCT) biometry in measuring ocular axial dimensions as well as imaging the intraocular lens (IOL) and posterior capsule in pseudophakic eyes.
methodsSixty-nine pseudophakic eyes of 69 subjects were enrolled in the study. The results of SW-OCT biometry implemented in the SD OCT device for posterior and anterior segment imaging (REVO NX, Optopol Technology) were compared with those obtained with the SS-OCT-based biometer IOLMaster 700 (Carl Zeiss Meditec). Differences in measurement values between the two biometers were determined using the paired t-test. Agreement was assessed through intraclass correlation coefficients (ICC) and Bland-Altman plots.
resultsThe correlation between measurements obtained with SW-OCT and SS-OCT was very high (ICC for: axial length (AL) = 1.000; anterior chamber depth (ACD) = 0.997; IOL thickness (IOL LT) = 0.997; central corneal thickness (CCT) = 0.987). The mean AL measurement difference was 0.003 ± 0.021 mm (the 95% LoA ranged from -0.04 to 0.05); the mean ACD difference was -0.009 ± 0.025 mm (95% LoA, -0.06 to 0.04); mean LT difference was 0.001 ± 0.021 mm (95% LoA, -0.04 to 0.04); and mean CCT difference was 1.4 ± 5.4 μm (95% LoA, -9 to 12).
conclusionThe study shows small, non-significant differences between the biometric measurements obtained with REVO NX SW-OCT and IOLMaster 700 SS-OCT in pseudophakic eyes. However, SW-OCT offered significantly lower ACD and LT measurement failure rates. With high-resolution imaging, SW-OCT enables accurate assessment of IOL position relative to the posterior capsule and visualization of capsular fibrosis.