Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi-leaf collimator quality assurance.

Simon K Goodall, Craig Norvill

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.5field-weighted citation impact, top 36% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 2 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors at 2 institutions in 2 countries.

Simon K GoodallGenesisCare, Wembley, Western Australia, Australia.ORCID https://orcid.org/0000-0003-3318-9703
Craig NorvillGenesisCare, Wembley, Western Australia, Australia.ORCID https://orcid.org/0000-0002-7724-5335
Genesis HealthCare · USThe University of Western Australia · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

For Elekta Agility linear accelerators, the iViewGT electronic portal imaging device (EPID) is positioned at a nominal X-Ray source-to-panel distance of 1600 mm. For display, image registration, and data processing purposes, the image pixels are scaled to spatial units at the treatment isocenter plane. This is achieved by applying a pixel scaling factor (PSF). During this investigation, the dependence of the PSF at cardinal gantry angles was determined along with the resulting effects on the multi-leaf collimator (MLC) quality assurance (QA) results for three linear accelerators (linacs). The PSF was found to vary by 0.0018-0.0022 mm/pixel during gantry rotation, which resulted in variations in the mean MLC reported error of up to 0.8 mm at 100 mm off-axis with the gantry rotated to 180°. Measurement and application of a gantry angle-specific PSF is a simple process that can be implemented to improve the accuracy of EPID-based MLC QA at cardinal gantry angles.

Indexed as

Electrical Equipment and SuppliesParticle AcceleratorsElectronicsHumansPhantoms, ImagingRotationDMLC QAEPIDMLC QA

Identifiers

PMID35666629
PMCPMC9278680
OpenAlexW4282918194

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.