Evidence map›Paper›PMID 35658905›Full record

ArticleJournal of orthopaedic surgery and research2022

A case-driven hypothesis for multi-stage crack growth mechanism in fourth-generation ceramic head fracture.

Stefano Lucchini, Massimiliano Baleani, Federico Giardina, Andrea Martelli, Francesco Castagnini, Barbara Bordini, Francesco Traina

Open access · goldAbstract read
In one paragraph

Article in Journal of orthopaedic surgery and research, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
2.0field-weighted citation impact, top 14% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 9 citations in OpenAlex.

  1. Article
  2. Article
  3. Global mapping of institutional and hospital-based (Level II-IV) arthroplasty registries: a scoping review.European journal of orthopaedic surgery & traumatology : orthopedie traumatologie · 2024
    Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 1 country.

Stefano LucchiniIRCCS Istituto Ortopedico Rizzoli, Ortopedia-Traumatologia e Chirurgia Protesica e dei Reimpianti d'Anca e Ginocchio, Bologna, Italy.
Massimiliano BaleaniIRCCS Istituto Ortopedico Rizzoli, Laboratorio di Tecnologia Medica, Bologna, Italy. massimiliano.baleani@ior.it.
Federico GiardinaIRCCS Istituto Ortopedico Rizzoli, Ortopedia-Traumatologia e Chirurgia Protesica e dei Reimpianti d'Anca e Ginocchio, Bologna, Italy.
Andrea MartelliIRCCS Istituto Ortopedico Rizzoli, Laboratorio di Tecnologia Medica, Bologna, Italy.
Francesco CastagniniIRCCS Istituto Ortopedico Rizzoli, Ortopedia-Traumatologia e Chirurgia Protesica e dei Reimpianti d'Anca e Ginocchio, Bologna, Italy.
Barbara BordiniIRCCS Istituto Ortopedico Rizzoli, Laboratorio di Tecnologia Medica, Bologna, Italy.
Francesco TrainaIRCCS Istituto Ortopedico Rizzoli, Ortopedia-Traumatologia e Chirurgia Protesica e dei Reimpianti d'Anca e Ginocchio, Bologna, Italy.
Istituto Ortopedico Rizzoli · ITIstituti di Ricovero e Cura a Carattere Scientifico · IT

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundCeramic bearings are used in total hip arthroplasty due to their excellent wear behaviour and biocompatibility. The major concern related to their use is material brittleness, which significantly impacts on the risk of fracture of ceramic components. Fracture toughness improvement has contributed to the decrease in fracture rate, at least of the prosthetic head. However, the root cause behind these rare events is not fully understood. This study evaluated head fracture occurrence in a sizeable cohort of patients with fourth-generation ceramic-on-ceramic implants and described the circumstances reported by patients in the rare cases of head fracture.

methodsThe clinical survivorship of 29,495 hip prostheses, with fourth-generation ceramic bearings, was determined using data from a joint replacement registry. The average follow-up period was 5.2 years (range 0.1-15.6). Retrieval analysis was performed in one case for which the ceramic components were available.

resultsClinical outcomes confirmed the extremely low fracture rate of fourth-generation ceramic heads: only two out of 29,495 heads fractured. The two fractures, both involving 36 mm heads, occurred without a concurrent or previous remarkable trauma. Considering the feature of the fractured head, a multi-stage crack growth mechanism has been hypothesized to occur following damage at the head-neck taper interface.

conclusionsSurgeons must continue to pay attention to the assembly of the femoral head: achieving a proper head seating on a clean taper is a prerequisite to decrease the risk of occurrence of any damage process within head-neck junction, which may cause high stress concentration at the contact surface, promoting crack nucleation and propagation even in toughened ceramics.

Indexed as

Arthroplasty, Replacement, HipFractures, BoneHip ProsthesisCeramicsFemur HeadHumansProsthesis DesignProsthesis FailureCeramic bearingCeramic head fractureFracture rateHip prosthesisMode of failure

Identifiers

PMID35658905
PMCPMC9164427
OpenAlexW4281873658

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.