Glow flow ionization mass spectrometry of small molecules. A comparison of a glow flow ionization source ('GlowFlow') with electrospray ionization and atmospheric pressure chemical ionization.
Article in Rapid communications in mass spectrometry : RCM, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 1 citations in OpenAlex.
- Glow flow ionization mass spectrometry of small molecules. A comparison of a glow flow ionization source ('GlowFlow') with electrospray ionization and atmospheric pressure chemical ionization.Rapid communications in mass spectrometry : RCM · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 2 countries.
Funding
Abstract
rationaleIonization by atmospheric pressure gas discharge has been employed for a long time in mass spectrometry. Inductively coupled plasma mass spectrometry is an exemplar, and widely used for elemental analysis. The technique has less uptake in organic mass spectrometry. We describe a simple source design that can be readily implemented in most atmospheric pressure ionization (API) systems and compare its performance with that of electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI).
methodsAn in-house designed helium gas discharge source (referred to as 'GlowFlow') was used on a Xevo G2-S time-of-flight mass spectrometer. The GlowFlow source was transferred to a compatible Xevo TQ-S triple-quadrupole mass spectrometer using an ultrahigh-performance liquid chromatograph inlet. Its performance was compared to that of Waters ESI and APCI sources.
resultsPreliminary results of GlowFlow on the Swansea instrument are presented to establish context and include analysis of low-molecular-mass polymers, benzoic acid and cinnamic acid. Comparison of performance on the Xevo TQ-S triple-quadrupole mass spectrometer involved three test mixtures. The method limits of detection (six-mix) for positive-ion GlowFlow source were between 0.03 and 10.00 pg with good linear response over two to four orders of magnitude and values of R
conclusionsThe current GlowFlow design is relatively simple to retrofit to most API systems due to its small size. The sensitivity of the GlowFlow design is typically an order of magnitude less than that of ESI in positive-ion mode, but similar in sensitivity in negative-ion mode and comparable to that of APCI.