In one paragraph

Article in Arthroscopy, sports medicine, and rehabilitation, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
0.7field-weighted citation impact, top 33% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 4 citations in OpenAlex.

  1. [Progress in surgical fixation methods for patellar osteochondral fractures following acute patellar dislocation].Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery · 2026
    Review
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 1 institution in 1 country.

Shizuka SasakiDepartment of Orthopaedic Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan.
Shohei YamauchiDepartment of Orthopaedic Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan.
Eiji SasakiDepartment of Orthopaedic Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan.
Yuka KimuraDepartment of Orthopaedic Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan.
Akane TakahashiDepartment of Orthopaedic Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan.
Yasuyuki IshibashiDepartment of Orthopaedic Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan.
Hirosaki University · JP

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: To compare the initial fixation strength of osteochondral fragment fixations using osteochondral plugs, bioabsorbable pins, and knotless suture anchors. Methods: Eighteen fresh-frozen immature (6 month old) porcine knees were used. An osteochondral fragment, cut from the articular surface of the medial femoral condyle to achieve a thickness of 5 mm, was used to mimic the unstable osteochondral fragment. It was fixed using three techniques, including two osteochondral plugs (osteochondral plug group), four full-threaded poly l-lactic acid pins (bioabsorbable pin group), and three suture anchors with a 2-0 tape (suture anchor group). Tensile loads at displacements of 1 and 2 mm and ultimate failure load were measured at a cross-head speed of 100 mm/min, and the variables of the three groups were compared statistically using a one-way ANOVA with Tukey's honestly significant difference test. Results: There was no significant difference in the tensile load to achieve 1-mm displacement. The load to achieve 2-mm displacement and the ultimate failure load were significantly greater in the suture anchor group than the osteochondral plug group and the bioabsorbable pin group. Conclusions: Single-pull destructive testing of a fixed articular osteochondral fragment with the force perpendicular to the articular surface, demonstrated no statistical difference in the tensile load to achieve 1-mm displacement, but the load to achieve 2-mm displacement was significantly greater for the three suture anchor-interlocking 2-0 tape constructs than the dual osteochondral plug fixation and the four bioabsorbable pin fixation constructs. Additionally, the three suture anchor-interlocking 2-0 tape construct's mean single-pull failure load was greater than other two fixation procedures. Clinical Relevance: To achieve osteochondral fragment union, sufficient fixation strength is critical. However, the initial fixation strength of osteochondral plugs, bioabsorbable pins, and knotless suture anchors for unstable osteochondral lesions remains unclear.