Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study.

Masahide Saito, Koji Ueda, Hidekazu Suzuki, Takafumi Komiyama, Kan Marino, Shinichi Aoki, Naoki Sano, Hiroshi Onishi

Open access · goldAbstract read
In one paragraph

Article in Journal of radiation research, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
2.1field-weighted citation impact, top 15% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 9 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 1 institution in 1 country.

Masahide SaitoDepartment of Radiology, University of Yamanashi, Yamanashi 409-3898, Japan.
Koji UedaDepartment of Radiology, University of Yamanashi, Yamanashi 409-3898, Japan.
Hidekazu SuzukiDepartment of Radiology, University of Yamanashi, Yamanashi 409-3898, Japan.
Takafumi KomiyamaDepartment of Radiology, University of Yamanashi, Yamanashi 409-3898, Japan.
Kan MarinoDepartment of Radiology, University of Yamanashi, Yamanashi 409-3898, Japan.
Shinichi AokiDepartment of Radiology, University of Yamanashi, Yamanashi 409-3898, Japan.
Naoki SanoDepartment of Radiology, University of Yamanashi, Yamanashi 409-3898, Japan.
Hiroshi OnishiDepartment of Radiology, University of Yamanashi, Yamanashi 409-3898, Japan.
University of Yamanashi · JP

Funding

Hamano Engineering
6 · The paper itself

Abstract

The purpose of this study is to evaluate the detection accuracy of a 3-dimensional (3D) body scanner, VOXELAN, in surface-guided radiotherapy (SGRT) of each part of the human body using a whole-body human phantom. We used A Resusci Anne was used as the whole-body phantom. The detection accuracy of VOXELAN in a radiotherapy treatment room with a linear accelerator (LINAC) was evaluated for two reference images: reconstruction of the planning computed tomography (CT) image (CT reference) and scanning by VOXELAN before the treatment (scan reference). The accuracy of the translational and rotational directions was verified for four treatment sites (open face shell, breast, abdomen, and arm), using the magnitude of the 6D robotic couch movement as the true value. Our results showed that the detection accuracy improved as the displacement from the reference position decreased for all the sites. Using the scan reference, the average accuracy of the translational and rotational axes was within 1.44 mm and 0.41°, respectively, for all sites except the arms. Similarly, using the CT reference, the average accuracy was within 2.45 mm and 1.35°, respectively. Additionally, it was difficult for both reference images to recognize misalignment of the arms. In conclusion we discovered that VOXELAN achieved a high detection accuracy for the head with an open face shell, chest, and abdomen, indicating that the system is useful in a clinical setting. However, it is necessary to pay attention to location matching for areas with few features, such as surface irregularities and potential errors, when the reference image is created from CT.

Indexed as

BrachytherapyRadiotherapy, Image-GuidedHumansParticle AcceleratorsPhantoms, ImagingRadiotherapy Planning, Computer-AssistedTomography, X-Ray Computedbreast cancerimage-guided radiotherapy (IGRT)optical imaging systemsurface-guided radiotherapy (SGRT)

Identifiers

PMID35467750
PMCPMC9124621
OpenAlexW4224435863

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.