Evidence map›Paper›PMID 35407908›Full record

ArticleMaterials (Basel, Switzerland)2022

Total Knee Replacement with an Uncemented Porous Tantalum Tibia Component: A Failure Analysis.

Samo K Fokter, Nenad Gubeljak, Esther Punzón-Quijorna, Primož Pelicon, Mitja Kelemen, Primož Vavpetič, Jožef Predan, Luka Ferlič, Igor Novak

Abstract read
In one paragraph

Article in Materials (Basel, Switzerland), 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 13 papers.

0numbers the graph read from it
0cells of the map it votes in
13citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

13 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Review
  6. Article
  7. Article
  8. Article
  9. Article
  10. Review
  11. Review
  12. Article
  13. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Samo K FokterDepartment of Orthopaedics, University Medical Centre Maribor, Ljubljanska 5, 2000 Maribor, Slovenia.ORCID 0000-0002-1902-8447
Nenad GubeljakFaculty of Mechanical Engineering, University of Maribor, Smetanova 17, 2000 Maribor, Slovenia.ORCID 0000-0002-3276-8431
Esther Punzón-QuijornaDepartment of Low and Medium Energy Physics F2, Jožef Stefan Institute, Jamova 39, 1000 Ljubljana, Slovenia.
Primož PeliconDepartment of Low and Medium Energy Physics F2, Jožef Stefan Institute, Jamova 39, 1000 Ljubljana, Slovenia.ORCID 0000-0003-2469-3342
Mitja KelemenDepartment of Low and Medium Energy Physics F2, Jožef Stefan Institute, Jamova 39, 1000 Ljubljana, Slovenia.
Primož VavpetičDepartment of Low and Medium Energy Physics F2, Jožef Stefan Institute, Jamova 39, 1000 Ljubljana, Slovenia.
Jožef PredanFaculty of Mechanical Engineering, University of Maribor, Smetanova 17, 2000 Maribor, Slovenia.
Luka FerličFaculty of Mechanical Engineering, University of Maribor, Smetanova 17, 2000 Maribor, Slovenia.
Igor NovakDepartment of Orthopaedics, University Medical Centre Maribor, Ljubljanska 5, 2000 Maribor, Slovenia.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Porous tantalum has been extensively used in orthopaedic surgery, including uncemented total knee arthroplasty (TKA). Favourable results were reported with earlier monobloc tibial components and the design evolved to modular implants. We aimed to analyse possible causes for extensive medial tibia bone loss, resulting in modular porous tantalum tibia baseplate fracture after primary TKA. Retrieved tissue samples were scanned with 3 MeV focused proton beam for Proton-Induced X-ray Emission (micro-PIXE) elemental analysis. Fractographic and microstructural analysis were performed by stereomicroscopy. A full 3D finite-element model was made for numerical analysis of stress-strain conditions of the tibial baseplate. Histological examination of tissue underneath the broken part of the tibial baseplate revealed dark-stained metal debris, which was confirmed by micro-PIXE to consist of tantalum and titanium. Fractographic analysis and tensile testing showed that the failure of the tibial baseplate fulfilled the criteria of a typical fatigue fracture. Microstructural analysis of the contact surface revealed signs of bone ingrowth in 22.5% of the surface only and was even less pronounced in the medial half of the tibial baseplate. Further studies are needed to confirm the responsibility of metal debris for an increased bone absorption leading to catastrophic tibial tray failure.

Indexed as

cementlesscementless baseplate fractureporous tantalumtibial componenttotal knee arthroplastytrabecular metaluncemented