Evidence map›Paper›PMID 35387136›Full record

ReviewJournal of arrhythmia2022

The utilization of atrial sensing dipole in single lead implantable cardioverter defibrillator for detection of new-onset atrial high-rate episodes or subclinical atrial fibrillation: A systematic review and meta-analysis.

Xuanming Pung, Daniel Zhihao Hong, Tzyy Yeou Ho, Xiayan Shen, Pei Ting Tan, Colin Yeo, Vern Hsen Tan

Abstract readReview
In one paragraph

Review in Journal of arrhythmia, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Article
  3. Ability to remotely monitor atrial high-rate episodes using a single-chamber implantable cardioverter-defibrillator with a floating atrial sensing dipole.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2023
    Article
  4. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Xuanming PungDepartment of Cardiology Changi General Hospital Singapore City Singapore.ORCID https://orcid.org/0000-0002-8261-131X
Daniel Zhihao HongYong Loo Lin School of Medicine National University of Singapore Singapore City Singapore.
Tzyy Yeou HoYong Loo Lin School of Medicine National University of Singapore Singapore City Singapore.
Xiayan ShenDepartment of Cardiology Changi General Hospital Singapore City Singapore.
Pei Ting TanHealth Services Research Changi General Hospital Singapore City Singapore.
Colin YeoDepartment of Cardiology Changi General Hospital Singapore City Singapore.
Vern Hsen TanDepartment of Cardiology Changi General Hospital Singapore City Singapore.ORCID https://orcid.org/0000-0003-0164-7925

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

This meta-analysis aims to evaluate the performance of atrial sensing dipole in single lead implantable cardioverter defibrillator (VDD-ICD) recipients in particular diagnosing new-onset atrial high-rate episodes (AHREs) defined as rate threshold of 200 beats per minute, or subclinical atrial fibrillation (SCAF) defined as device-detected AF without symptoms. We comprehensively searched PubMed, Embase, and ClinicalTrials.gov. Studies comparing contemporary single- and dual-chamber ICD (VVI-/DDD-ICD) versus VDD-ICD were included. Restricted maximum likelihood method for random effect model and Mantel-Haenszel method for fixed effect model were used to estimate the effect size of new-onset AHREs, or SCAF detection in each group. Three prospective studies were identified and total of 991 participants were included. There were 330 (33.3%) in VDD-ICD and 661 (66.7%) in VVI-/DDD-ICD. Most (78%) participants were men. Median follow-up was from 365 days to 847 days. VDD-ICD has a higher likelihood of detecting AHREs or SCAF as compared to VVI-/DDD-ICD [(OR random effect : 2.6; 95% CI: 1.2, 5.8;

Indexed as

atrial high rate episodesatrial sensing dipoleimplantable cardiac defibrillatorsubclinical atrial fibrillation

Identifiers

PMID35387136
PMCPMC8977580

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked