Evidence map›Paper›PMID 35371384›Full record

ArticleKansas journal of medicine2022

Stacked 1/3 Tubular Plates for Fixation of Pediatric Forearm Fractures: A Biomechanical Study.

Benjamin J Cooper, Alexander C Wendling, Sharon R Isaacs-Pullins, Joel White, Steven M Hollenbeck

Abstract read
In one paragraph

Article in Kansas journal of medicine, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Mechanical comparison of different plate-and-screw configurations for repairing forearm diaphysis fractures: a systematic review.European journal of trauma and emergency surgery : official publication of the European Trauma Society · 2026
    Pooled it
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Benjamin J CooperUniversity of Kansas School of Medicine-Wichita, Wichita, KS.
Alexander C WendlingUniversity of Kansas School of Medicine-Wichita, Wichita, KS.
Sharon R Isaacs-PullinsUniversity of Kansas School of Medicine-Wichita, Wichita, KS.
Joel WhiteNational Institute for Aviation Research, Wichita State University, Wichita, KS.
Steven M HollenbeckUniversity of Kansas School of Medicine-Wichita, Wichita, KS.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Introduction: Among operatively treated pediatric forearm fractures, many different fixation constructs are described. The goal of this study was to define the biomechanical properties of a double stacked 1/3 tubular plate construct used by the senior author for some fractures and to review available literature regarding the use of stacked plates. Methods: Biomechanical testing was performed by 4-point bending of three different plate constructs: 1/3 tubular plate, stacked 1/3 tubular plates, and 2.7 mm LC-DCP plate. Five test specimens were evaluated for each of the three plate constructs. From stress-strain curves, flexural stiffness (N/mm), force to cause plastic deformation (N), and force to cause 10° bend (N) were calculated and compared using standard t-test statistics. Results: Key outcome parameter means (± SD) for the three plate constructs (1/3 tubular plate, stacked 1/3 tubular plates, and 2.7 mm LC-DCP plate) were reported respectively as follows: flexural stiffness (55.4 ± 3.5 N/mm, 131.7 ± 3.5 N/mm, 113.3 ± 12.1 N/mm), force to cause plastic deformation (113.6 ± 11.0 N, 242.1 ± 13.0 N, 192.2 ± 17.9 N), and force to cause a 10° bend (140.0 ± 8.4 N, 299.4 ± 14.1 N, 265.5 ± 21.2 N). Mean values of all three measures were significantly larger for the stacked 1/3 tubular plates than for the other plate constructs. Conclusions: The stacked 1/3 tubular plate construct was biomechanically superior to the other plate constructs tested. Stacked plating significantly improved stiffness of the fracture fixation construct supporting the use of this technique in selected trauma cases.

Indexed as

biomechanicsbone platesequipment designfracture fixationmaterials testing

Identifiers

PMID35371384
PMCPMC8942585

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.