Evidence map›Paper›PMID 35347340›Full record

ArticleActa orthopaedica2022

Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation.

Peter Bo Jørgensen, Bart L Kaptein, Kjeld Søballe, Stig S Jakobsen, Maiken Stilling

Open access · goldAbstract read
In one paragraph

Article in Acta orthopaedica, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
1.7field-weighted citation impact, top 16% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 8 citations in OpenAlex.

  1. Article
  2. Article
  3. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 2 institutions in 2 countries.

Peter Bo JørgensenAutoRSA Research Group, Orthopeadic Research Unit Aarhus University Hospital, Aarhus. pbjr@clin.au.dk.
Bart L KapteinBiomechanics and Imaging Group, Department of Orthopaedics, Leiden University Medical Center, Leiden, The Netherlands. b.l.kaptein@lumc.nl.
Kjeld SøballeAutoRSA Research Group, Orthopaedic Research Unit, Aarhus University Hospital, Aarhus, Denmark; Biomechanics and Imaging Group, Department of Orthopaedics, Leiden University Medical Center, Leiden, The Netherlands. kjeld@soballe.com.
Stig S JakobsenDepartment of Orthopaedic Surgery, Aarhus University Hospital, Aarhus N, Denmark. stigjako@rm.dk.
Maiken StillingAutoRSA Research Group, Orthopaedic Research Unit, Aarhus University Hospital, Aarhus, Denmark; Department of Orthopaedic Surgery, Aarhus University Hospital, Aarhus N, Denmark. maiken.stilling@clin.au.dk.
Aarhus University Hospital · DKLeiden University Medical Center · NL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

background and purposeDual-mobility hip arthroplasty utilizes a freely rotating polyethylene acetabular liner to protect against dislocation. As liner motion has not been confirmed in vivo, we undertook this using dynamic radiostereometry (RSA). PATIENTS AND

methods6 patients with Anatomical Dual Mobility acetabular components were included. Markers were implanted in the liners using a drill guide. Static RSA recordings and patient-reported outcome measures were obtained postoperatively and at 1-year follow-up. Dynamic RSA recordings were obtained at 1-year follow-up during passive hip movement: abduction/external rotation, adduction/internal rotation (modified FABER-FADIR), to end-range and at 45° hip flexion. Liner and neck movements were described as anteversion, inclination, and rotation.

resultsLiner movement during modified FABER-FADIR was detected in 12 of 16 patients. Median (range) absolute liner movements were: anteversion 10° (5-20), inclination 6° (2-12), and rotation 11° (5-48) relative to the cup. Median absolute change in the resulting liner/neck angle (small articulation) was 28° (12-46) and in liner/cup angle (larger articulation) was 6° (4-21). Static RSA showed changes in median liner anteversion from 7° (-12 to 23) postoperatively to 10° (-3 to 16) at 1-year follow-up and inclination from 42° (35-66) postoperatively to 59° (46-80) at 1-year follow-up. Liner/neck contact was associated with high initial liner anteversion (p = 0.01).

interpretationThe polyethylene liner moves over time. 1 year after surgery the liner can move with or without liner/neck contact. The majority of movement is in the smaller articulation between head and liner.

Indexed as

Arthroplasty, Replacement, HipHip ProsthesisHumansPolyethyleneRadiostereometric AnalysisRange of Motion, ArticularPolyethylene

Identifiers

PMID35347340
PMCPMC8958851
OpenAlexW4220768099

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.