Evidence map›Paper›PMID 35338583›Full record

ArticleJournal of applied clinical medical physics2022

Evaluation of a proprietary software application for motion monitoring during stereotactic paraspinal treatment.

Qiyong Fan, Hai Pham, Pengpeng Zhang, Xiang Li, Tianfang Li

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
1.5field-weighted citation impact, top 19% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 7 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 1 institution in 1 country.

Qiyong FanDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
Hai PhamDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
Pengpeng ZhangDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
Xiang LiDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
Tianfang LiDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
Memorial Sloan Kettering Cancer Center · US

Funding

X-RAY CRYSTALLOGRAPHYP30CA008748 · NCI · SLOAN-KETTERING INSTITUTE FOR CANCER RES · PI Nicholas D Socci · 1985 to 2026
$347.4M
NCI NIH HHS P30 CA008748
6 · The paper itself

Abstract

purposeStereotactic paraspinal treatment has become increasingly popular due to its favorable clinical outcome. An often-overlooked factor that compromises the effectiveness of such treatment is the patients' involuntary intrafractional motion. This work introduces and validates a proprietary software application that quantifies such motion for accurate patient monitoring during treatment.

methodsThe software uses a separate full-trajectory cone-beam computed tomography (CBCT) after daily patient setup to establish reference projections. Once treatment starts, the software grabs the intrafraction motion review (IMR) image acquired by TrueBeam via the Varian iTools Capture software and compares it against the corresponding reference projection to instantly determine the 2D shifts of the vertebrae being monitored using the classical downhill simplex optimization method. To evaluate its performance, an anthropomorphic phantom was shifted 0, 0.6, 1.2, 1.8, 2.4, 3.0, and 5 mm in three orthogonal directions, immediately after the full-trajectory CBCT but prior to treatment. Depending on the scenario of shift, a nine-field fixed gantry intensity-modulated radiation therapy (IMRT) plan and/or a four partial-posterior-arcs volume-modulated radiation therapy (VMAT) plan were delivered. For the IMRT plan, three IMR images were acquired sequentially every 200 monitor units (MU) at each treatment angle. For the VMAT plan, one IMR image was acquired every 15° of each arc. For each IMR image, the software-reported 2D shift was compared with the ground truth. Certain tests were repeated with 1°, 2°, and 3° of rotation, pitch, and roll, respectively. Some of these tests were also repeated independently on separate days.

resultsBased on the group of tests that involved only the IMRT delivery, the maximum standard deviation of the software-reported shifts for each set of three IMR images was 0.16 mm, with 95th percentile at 0.02 mm. For translational shift, the maximum registration error was 0.44 mm, with 95th percentile at 0.23 mm. Left unaccounted for, rotation and pitch degraded the registration accuracy mainly in the longitudinal direction, while roll degraded it mainly in the lateral direction. The degradation of registration accuracy is positively related to the degree of rotation, pitch, and roll. The maximum registration errors under 3° rotation, pitch, and roll were 2.97, 1.44, 2.72 mm, respectively. Based on the group of tests that compared IMRT delivery with VMAT delivery, the registration errors slightly increased as magnitude of shifts increased; however, they were well under the 0.5-mm threshold. No significant differences in registration errors were observed between IMRT and VMAT deliveries. In addition, the variation in registration errors among different days was limited for both IMRT and VMAT deliveries.

conclusionsOur proprietary software has high repeatability, both intrafractionally and interfractionally, and high accuracy in registering IMR images with the reference projections for motion monitoring, regardless of the magnitude of shifts or treatment delivery technique. Rotation, pitch, and roll degrade registration accuracy and need to be accounted for in the future work.

Indexed as

Radiotherapy, Intensity-ModulatedCone-Beam Computed TomographyHumansPhantoms, ImagingRadiotherapy Planning, Computer-AssistedRotationSoftwareIMRKV imagingmarkerless spine trackingparaspinal SBRT