Evidence map›Paper›PMID 35336379›Full record

ArticleSensors (Basel, Switzerland)2022

Utility of the Novel MediPost Mobile Posturography Device in the Assessment of Patients with a Unilateral Vestibular Disorder.

Oskar Rosiak, Anna Gawronska, Magdalena Janc, Pawel Marciniak, Rafal Kotas, Ewa Zamyslowska-Szmytke, Magdalena Jozefowicz-Korczynska

Abstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed.

  1. Review
  2. Article
  3. Inertial Measurement Unit-Based Romberg Test for Assessing Adults With Vestibular Hypofunction.IEEE journal of translational engineering in health and medicine · 2024
    Article
  4. Article
  5. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Oskar RosiakBalance Disorders Unit, Department of Otolaryngology, Medical University of Lodz, The Norbert Barlicki Memorial Teaching Hospital, 90-153 Lodz, Poland.ORCID 0000-0002-5326-4777
Anna GawronskaBalance Disorders Unit, Department of Otolaryngology, Medical University of Lodz, The Norbert Barlicki Memorial Teaching Hospital, 90-153 Lodz, Poland.
Magdalena JancAudiology and Phoniatrics Clinic, Nofer Institute of Occupational Medicine, 91-348 Lodz, Poland.ORCID 0000-0003-4527-6513
Pawel MarciniakDepartment of Microelectronics and Computer Science, Lodz University of Technology, 90-924 Lodz, Poland.ORCID 0000-0001-6400-0104
Rafal KotasDepartment of Microelectronics and Computer Science, Lodz University of Technology, 90-924 Lodz, Poland.ORCID 0000-0003-0776-9173
Ewa Zamyslowska-SzmytkeAudiology and Phoniatrics Clinic, Nofer Institute of Occupational Medicine, 91-348 Lodz, Poland.
Magdalena Jozefowicz-KorczynskaBalance Disorders Unit, Department of Otolaryngology, Medical University of Lodz, The Norbert Barlicki Memorial Teaching Hospital, 90-153 Lodz, Poland.

Funding

National Center for Research and Development STRATEGMED 2/266299/19NCBR/201
6 · The paper itself

Abstract

Balance disorders are a growing problem worldwide. Thus, there is an increasing need to provide an inexpensive and feasible alternative to standard posturographic platforms (SP) used for the assessment of balance and to provide a possible solution for telemonitoring of patients. A novel mobile posturography (MP) MediPost device was developed to address these issues. This prospective study used a Modified Clinical Test of Sensory Interaction on Balance to evaluate healthy individuals and patients with a unilateral vestibular disorder through SP and MP simultaneously. The control group included 65 healthy volunteers, while the study group included 38 patients diagnosed with a unilateral vestibular deficit. The angular velocity values obtained from both methods were compared by intraclass correlation coefficients (ICC) and Bland−Altman plot analysis. Diagnostic capabilities were measured in terms of sensitivity and specificity. The ICC between the two methods for conditions 2−4 was indicative of excellent reliability, with the ICC > 0.9 (p < 0.001), except for Condition 1 (standing stance, eyes open) ICC = 0.685, p < 0.001, which is indicative of moderate reliability. ROC curve analysis of angular velocity for condition 4 represents the most accurate differentiating factor with AUC values of 0.939 for SP and 0.953 for MP. This condition also reported the highest sensitivity, specificity, PPV, and NPV values with 86.4%, 87.7%, 80%, and 90.5% for SP, and 92.1%, 84.6%, 77.8%, and 94.8% for MP, respectively. The newly developed MediPost device has high sensitivity and specificity in distinguishing between healthy individuals and patients with a unilateral vestibular deficit.

Indexed as

Postural BalanceVestibular DiseasesHumansPhysical Therapy ModalitiesProspective StudiesReproducibility of Resultsbalance disordersBland–Altmannbody-worn sensorsintraclass correlation coefficientsmobile posturographyvertigo

Identifiers

PMID35336379
PMCPMC8953915

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.