ArticleEJNMMI physics2022

Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard.

Nidhal Kahlaoui, Thibaut Naninck, Roger Le Grand, Catherine Chapon

Open access · goldAbstract read
In one paragraph

Article in EJNMMI physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.2field-weighted citation impact, top 52% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 2 citations in OpenAlex.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 1 institution in 1 country.

Nidhal KahlaouiInserm, CEA, Center for Immunology of Viral, Auto-immune, Hematological and Bacterial Diseases (IMVA-HB/IDMIT), Fontenay-aux-Roses & Le Kremlin-Bicêtre, Université Paris-Saclay, Paris, France.
Thibaut NaninckInserm, CEA, Center for Immunology of Viral, Auto-immune, Hematological and Bacterial Diseases (IMVA-HB/IDMIT), Fontenay-aux-Roses & Le Kremlin-Bicêtre, Université Paris-Saclay, Paris, France.
Roger Le GrandInserm, CEA, Center for Immunology of Viral, Auto-immune, Hematological and Bacterial Diseases (IMVA-HB/IDMIT), Fontenay-aux-Roses & Le Kremlin-Bicêtre, Université Paris-Saclay, Paris, France.
Catherine ChaponInserm, CEA, Center for Immunology of Viral, Auto-immune, Hematological and Bacterial Diseases (IMVA-HB/IDMIT), Fontenay-aux-Roses & Le Kremlin-Bicêtre, Université Paris-Saclay, Paris, France. catherine.chapon@cea.fr.ORCID http://orcid.org/0000-0002-3210-8468
Inserm · FR

Funding

Agence Nationale de la Recherche ANR-11-INBS-0008
6 · The paper itself

Abstract

introductionA Vereos PET/CT device was adapted to be compatible with the experimentation in large animals within BSL-3 environment. The aim of this study was to investigate the impact of this modification on the performance according to NEMA NU2-2012 standard.

methodsSpatial resolution, sensitivity, count rate performance, accuracies of corrections and image quality were assessed using the NEMA NU2-2012 standards before and after installation of a transparent poly-methyl methacrylate tube of 8 mm thickness, 680 mm diameter and 2800 mm long inside the tunnel of the system. In addition, CT performance tests were performed according to manufacturer standard procedure.

resultsAlthough the presence of the tube led to a slight decrease in sensitivity, performance measurements were in accordance with manufacturer preconisation ranges and comparable to previous performance published data.

conclusionModifications of Vereos PET/CT system allowing its use in BSL-3 conditions did not affect significantly its performance according to NEMA NU2-2012 standard. KEY POINTS: Question. Does a BSL-3 compatible modification alter Philips Vereos PET/CT performances according to NEMA NU2-2012 standards? Pertinent findings. Our Vereos PET/CT system was modified by a wall separating BSL-1 and BSL-3 sides and an 8 mm thickness PMMA tube inserted into the bore of the camera in order to extend the BSL-3 containment along the bed movement. The performances of our modified system according to NEMA NU2-2012 standards were not significantly impacted by the modifications and were in accordance with the values prescribed by the manufacturer. Implications for patients care. Our clinical PET/CT device was modified for human infectious diseases studies in Non-Human Primates. This unusual set up may then provide truly transposable data from preclinical studies into clinical application in infected patients.

Indexed as

BSL-3 conditionsInfectious diseasesNEMA NU2-2012 standardPET/CT

Identifiers

PMID35316846
PMCPMC8938744
OpenAlexW4220692733

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.