ArticleFrontiers in physiology2022

Low Ambient Temperature Exposition Impairs the Accuracy of a Non-invasive Heat-Flux Thermometer.

Michela Masè, Andreas Werner, Gabriel Putzer, Giovanni Avancini, Marika Falla, Hermann Brugger, Alessandro Micarelli, Giacomo Strapazzon

Abstract read
In one paragraph

Article in Frontiers in physiology, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Michela MasèInstitute of Mountain Emergency Medicine, Eurac Research, Bolzano, Italy.
Andreas WernerInstitute of Physiology, Center for Space Medicine and Extreme Environments Berlin, Charité - Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, Berlin, Germany.
Gabriel PutzerDepartment of Anaesthesiology and Intensive Care Medicine, Medical University of Innsbruck, Innsbruck, Austria.
Giovanni AvanciniDepartment of Anaesthesia and Intensive Care, Santa Chiara Hospital, Trento, Italy.
Marika FallaInstitute of Mountain Emergency Medicine, Eurac Research, Bolzano, Italy.
Hermann BruggerInstitute of Mountain Emergency Medicine, Eurac Research, Bolzano, Italy.
Alessandro MicarelliInstitute of Mountain Emergency Medicine, Eurac Research, Bolzano, Italy.
Giacomo StrapazzonInstitute of Mountain Emergency Medicine, Eurac Research, Bolzano, Italy.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Indirect core body temperature (CBT) monitoring from skin sensors is gaining attention for in-field applications thanks to non-invasivity, portability, and easy probe positioning. Among skin sensors, heat-flux devices, such as the so-called Double Sensor (DS), have demonstrated reliability under various experimental and clinical conditions. Still, their accuracy at low ambient temperatures is unknown. In this randomized cross-over trial, we tested the effects of cold temperature exposition on DS performance in tracking CBT. Methods: Twenty-one participants were exposed to a warm (23.2 ± 0.4°C) and cold (-18.7 ± 1.0°C) room condition for 10 min, following a randomized cross-over design. The accuracy of the DS to estimate CBT in both settings was assessed by quantitative comparison with esophageal (reference) and tympanic (comparator) thermometers, using Bland-Altman and correlation analyses (Pearson's correlation coefficient, Results: In the warm room setting, the DS showed a moderate agreement with the esophageal sensor [bias = 0.09 (-1.51; 1.69) °C, Conclusion: DS accuracy is influenced by environmental conditions and previous exposure to cold settings. These results suggest the present inadequacy of the DS device for in-field applications in low-temperature environments and advocate further technological advancements and proper sensor insulation to improve performance in these conditions.

Indexed as

body temperaturecold environmentshypothermianon-invasive devicesphysiological monitoringwearables

Identifiers

PMID35309078
PMCPMC8931521

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.