Alinity hq platelet count is not impacted by severe microcytosis.
Article in Journal of clinical laboratory analysis, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Concordance between platelet counts by impedance and optical technique during microcytic anemia: towards a threshold value of the mean corpuscular volume.La Tunisie medicale · 2024Article
- Alinity hq platelet count is not impacted by severe microcytosis.Journal of clinical laboratory analysis · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
backgroundImpedance technology has been shown to overestimate platelet (PLT) count in samples with microcytes, while the optical-fluorescence PLT count (PLT-F) by Sysmex has been suggested to be unaffected by microcytosis. The Abbott Alinity hq analyzer employs multi-dimensional optical PLT counting. Our goal was to assess the accuracy of this technology in microcytic samples.
methodsPlatelet measurements were performed by Alinity hq and the impedance (PLT-I) and PLT-F methods on a Sysmex XN-3000 analyzer on 464 samples. PLT concentration range was 6.56-947 × 10
resultsAlinity hq PLT count showed excellent agreement with PLT-F (r = 1.00). Sysmex PLT-I data showed somewhat weaker correlation with both PLT-F and Alinity hq (r = 0.98). Increasing bias between Sysmex PLT-I and PLT-F was seen with decreasing MCV values, with mean bias of 35.2 × 10
conclusionThe Sysmex PLT-I method overestimated the PLT count in samples with severe microcytosis. Alinity hq provided PLT counts and PLT and RBC indices that were not impacted by microcytosis.