ArticleStrahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al]2022
Local dose rate effects in implantable cardioverter-defibrillators with flattening filter free and flattened photon radiation.
Article in Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al], 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers, 2 of them syntheses that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 2 syntheses or guidelines pooled it.
- Systematic review of physical and dosimetric criteria for managing CIEDs in radiotherapy: A medical physicist's perspective.Journal of applied clinical medical physics · 2026Pooled it
- A systematic review and meta-analysis on oncological radiotherapy in patients with a cardiac implantable electronic device: Prevalence and predictors of device malfunction in 3121 patients.European journal of clinical investigation · 2023Pooled it
- Neutron dose from a 6-MV X-ray beam in radiotherapy.Radiological physics and technology · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeIn the beam penumbra of stereotactic body radiotherapy volumes, dose rate effects in implantable cardioverter-defibrillators (ICDs) may be the predominant cause for failures in the absence of neutron-generating photon energies. We investigate such dose rate effects in ICDs and provide evidence for safe use of lung tumor stereotactic radioablation with flattening filter free (FFF) and flattened 6 Megavolt (MV) beams in ICD-bearing patients.
methodsSixty-two ICDs were subjected to scatter radiation in 1.0, 2.5, and 7.0 cm distance to 100 Gy within a 5 × 5 cm
resultsWith 6 MV FFF beams, ICD errors occurred at distances of 1.0 cm (LDR 46.8 cGy/min; maximum ICD dose 3.4 Gy) and 2.5 cm (LDR 15.6 cGy/min; 1.1 Gy). With 6 MV FLAT beams, ICD errors occurred only at 1 cm distance (LDR 16.8 cGy/min; 3.9 Gy). No errors occurred at an LDR below 7 cGy/min, translating to a safe distance of 2.5 cm (1.5 Gy) in flattened and 7 cm (0.4 Gy) in 6 MV FFF beams.
conclusionA LDR in ICDs larger than 7 cGy/min may cause ICD malfunction. At identical LDR, differences between 6 MV FFF and 6 MV FLAT beams do not yield different rates of malfunction. The dominant reason for ICD failures could be the LDR and not the total dose to the ICD. For most stereotactic treatments, it is recommended to generate a planning risk volume around the ICD in which LDR larger than 7 cGy/min are avoided.