ArticleSurgical endoscopy2022

Analyzing material changes consistent with degradation of explanted polymeric hernia mesh related to clinical characteristics.

Xinyue Lu, Melinda Harman, B Todd Heniford, Vedra Augenstein, Brittney McIver, William Bridges

Open access · greenAbstract read
In one paragraph

Article in Surgical endoscopy, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
1.1field-weighted citation impact, top 24% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 6 citations in OpenAlex.

  1. Macroscopic, radiologic, and microscopic analysis of explanted abdominal wall prosthesis.Hernia : the journal of hernias and abdominal wall surgery · 2026
    Article
  2. Article
  3. Article
  4. Characterization in respect to degradation of titanium-coated polypropylene surgical mesh explanted from humans.Journal of biomedical materials research. Part B, Applied biomaterials · 2023
    Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 2 institutions in 1 country.

Xinyue LuDepartment of Bioengineering, Clemson University, Clemson, SC, 29634, USA.
Melinda HarmanDepartment of Bioengineering, Clemson University, Clemson, SC, 29634, USA. harman2@clemson.edu.ORCID http://orcid.org/0000-0001-7164-6552
B Todd HenifordGastrointestinal and Minimally Invasive Surgery, Carolinas Medical Center, Charlotte, NC, 28204, USA.
Vedra AugensteinGastrointestinal and Minimally Invasive Surgery, Carolinas Medical Center, Charlotte, NC, 28204, USA.
Brittney McIverDepartment of Bioengineering, Clemson University, Clemson, SC, 29634, USA.
William BridgesDepartment of Mathematical Sciences, Clemson University, Clemson, SC, 29634, USA.
Clemson University · USCarolinas Medical Center · US

Funding

Tissue Structural and Neural Remodeling in Human Sacroiliac JointP20GM121342 · NIGMS · CLEMSON UNIVERSITY · PI Hai Yao · 2018 to 2026
$24.7M
Development of a Novel Bio-Adhesive Mesh System for Abdominal Hernia RepairR41DK120168 · NIDDK · CIRCA BIOSCIENCE, LLC · PI CHAMPAIGNE, KEVIN D · 2019 to 2019
$224k
NIDDK NIH HHS R41 DK120168NIGMS NIH HHS P20 GM121342NIH HHS P20GM121342
6 · The paper itself

Abstract

backgroundProposed mechanisms that potentially contribute to polypropylene mesh degradation after in vivo exposure include oxidizing species and mechanical strains induced by normal healing, tissue integration, muscle contraction, and the immediate and chronic inflammatory responses.

methodsThis study explores these potential degradation mechanisms using 63 mesh implants retrieved from patients after a median implantation time of 24 months following hernia repair surgery (mesh explants) and analysis of multivariate associations between the material changes and clinical characteristics. Specifically, polypropylene mesh degradation was characterized in terms of material changes in surface oxidation, crystallinity and mechanical properties, and clinical characteristics included mesh placement location, medical history and mesh selection.

resultsCompared to pristine control samples, subsets of mesh explants had evidence of surface oxidation, altered crystallinity, or changed mechanical properties. Using multivariate statistical approach to control for clinical characteristics, infection was a significant factor affecting changes in mesh stiffness and mesh class was a significant factor affecting polypropylene crystallinity changes.

conclusionsHighly variable in vivo conditions expose mesh to mechanisms that alter clinical outcomes and potentially contribute to mesh degradation. These PP mesh explants after 0.5 to 13 years in vivo had measurable changes in surface chemistry, crystallinity and mechanical properties, with significant trends associated with factors of mesh placement, mesh class, and infection.

Indexed as

PolypropylenesSurgical MeshHerniaHumansMaterials TestingProstheses and ImplantsPolypropylenesDegradationHerniaMechanical strainMeshPolypropyleneReactive oxygen species (ROS)

Identifiers

PMID35257210
PMCPMC10851311
OpenAlexW4220661261

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.