Evidence map›Paper›PMID 35255846›Full record

ArticleBMC pediatrics2022

Dose error reduction software in medication safety risk management - optimising the smart infusion pump dosing limits in neonatal intensive care unit prior to implementation.

Sini Kuitunen, Krista Kärkkäinen, Carita Linden-Lahti, Lotta Schepel, Anna-Riia Holmström

Abstract read
In one paragraph

Article in BMC pediatrics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Closed-Loop Medication Management with an Electronic Health Record System in U.S. and Finnish Hospitals.International journal of environmental research and public health · 2023
    Review
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Sini KuitunenHUS Pharmacy and HUS Children and Adolescents, Helsinki University Hospital, Helsinki, Finland. sini.kuitunen@hus.fi.
Krista KärkkäinenDivision of Pharmacology and Pharmacotherapy, Faculty of Pharmacy, University of Helsinki, Helsinki, Finland.
Carita Linden-LahtiDivision of Pharmacology and Pharmacotherapy, Faculty of Pharmacy, University of Helsinki, Helsinki, Finland.
Lotta SchepelHUS Joint Resources and HUS Pharmacy, Helsinki University Hospital, Helsinki, Finland.
Anna-Riia HolmströmDivision of Pharmacology and Pharmacotherapy, Faculty of Pharmacy, University of Helsinki, Helsinki, Finland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundSmart infusion pumps with dose error reduction software can be used to prevent harmful medication errors. The aim of this study was to develop a method for defining and assessing optimal dosing limits in a neonatal intensive care unit's smart infusion pump drug library by using simulation-type test cases developed based on medication error reports.

methodsThis mixed-methods study applied both qualitative and quantitative methods. First, wrong infusion rate-related medication errors reported in the neonatal intensive care unit during 2018-2019 were explored by quantitative descriptive analysis and qualitative content analysis to identify the error mechanisms. The researchers developed simulation-type test cases with potential errors, and a literature-based calculation formula was used to set upper soft limits to the drug library. The limits were evaluated by conducting programming of pumps without errors and with potential errors for two imaginary test patients (1 kg and 3.5 kg).

resultsOf all medication errors reported in the neonatal intensive care unit, 3.5% (n = 21/601) involved an error or near-miss related to wrong infusion rate. Based on the identified error mechanisms, 2-, 5-, and 10-fold infusion rates, as well as mix-ups between infusion rates of different drugs, were established as test cases. When conducting the pump programming for the test cases (n = 226), no alerts were triggered with infusion rates responding to the usual dosages (n = 32). 73% (n = 70/96) of the erroneous 2-, 5-, and 10-fold infusion rates caused an alert. Mix-ups between infusion rates triggered an alert only in 24% (n = 24/98) of the test cases.

conclusionsSimulation-type test cases can be applied to assess the appropriateness of dosing limits within the neonatal intensive care unit's drug library. In developing the test cases, combining hospital's medication error data to other prospective data collection methods is recommended to gain a comprehensive understanding on mechanisms of wrong infusion rate errors. After drug library implementation, the alert log data and drug library compliance should be studied to verify suitability of dosing limits.

Indexed as

Infusion PumpsIntensive Care Units, NeonatalHumansInfant, NewbornInfusions, IntravenousRisk ManagementSoftwareDose error reduction softwareError reportingHigh-alert medicationMedication errorMedication safetyNeonatal intensive care unitProactive risk managementSmart infusion pump