A wearable cardioverter defibrillator with a low false alarm rate.
Article in Journal of cardiovascular electrophysiology, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT03887052 (ASSURE WCD Clinical Evaluation - Detection and Safety Study), which is not on this map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
ASSURE WCD Clinical Evaluation - Detection and Safety Study (ACE-DETECT)
Who cites it
9 citing papers in PubMed.
- First human safety and effectiveness study of defibrillation with a novel patch wearable cardioverter-defibrillator.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2024Trial
- Wearable Cardioverter-Defibrillator in Patients at Transient Risk of Sudden Cardiac Death: State of the Art and Contemporary Clinical Evidence.Diagnostics (Basel, Switzerland) · 2026Review
- Wearable Cardioverter-Defibrillator in Adult Congenital Heart Disease: A Bridge to Decision, Recovery, and Advanced Therapies.Heart failure reviews · 2026Review
- Variabiliy in Prescribing Patterns of Wearable Cardiac Defibrillators: Current Data, Guidelines, Challenges, and Controversies.Current cardiology reports · 2025Review
- A review on wearable cardioverter-defibrillators in clinical practice for sudden cardiac death prevention and heart failure management.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2025Review
- False Alarms in Wearable Cardioverter Defibrillators-A Relevant Issue or an Insignificant Observation.Journal of clinical medicine · 2024Article
- The Role of the Wearable Defibrillator in Heart Failure.Current heart failure reports · 2024Review
- The wearable cardioverter-defibrillator-Improving comfort and reaching towards noise immunity.Journal of cardiovascular electrophysiology · 2022Article
- A wearable cardioverter defibrillator with a low false alarm rate.Journal of cardiovascular electrophysiology · 2022Article
Corrections and comments
- Commented on by
Authors and funding
17 authors.
Funding
Abstract
introductionA wearable cardioverter defibrillator (WCD) is indicated in appropriate patients to reduce the risk for sudden cardiac death. Challenges for patients wearing a WCD have been frequent false shock alarms primarily due to electrocardiogram noise and wear discomfort. The objective of this study was to test a contemporary WCD designed for reduced false shock alarms and improved comfort.
methodsOne hundred and thirty patients with left ventricular ejection fraction ≤40% and an active implantable cardioverter defibrillator (ICD) were fitted with the ASSURE WCD (Kestra Medical Technologies) and followed for 30 days. WCD detection was enabled and shock alarm markers recorded, but shocks and shock alarms were disabled. All WCD episodes and ICD ventricular tachycardia/ventricular fibrillation (VT/VF) episodes were adjudicated. The primary endpoint was the false-positive shock alarm rate with a performance goal of 1 every 3.4 days (0.29 per patient-day).
resultsOf 163 WCD episodes, 4 were VT/VF and 159 non-VT/VF (121 rhythms with noise, 32 uncertain with noise, 6 atrial flutter without noise). Only three false-positive shock alarm markers were recorded; one false-positive shock alarm every 1333 patient-days (0.00075 per patient-day, 95% confidence interval: 0.00015-0.00361; p < .001). No ICD recorded VT/VF episodes meeting WCD detection criteria (≥170 bpm for ≥20 s) were missed by the WCD during 3501 patient-days of use. Median wear was 31.0 days (interquartile range [IQR] 2.0) and median daily use 23.0 h (IQR 1.7). Adverse events were mostly mild: skin irritation (19.4%) and musculoskeletal discomfort (8.5%).
conclusionThe ASSURE WCD demonstrated a low false-positive shock alarm rate, low patient-reported discomfort, and no serious adverse events.