ArticleJournal of the American College of Emergency Physicians open2022
Development and efficacy testing of a portable negative pressure enclosure for airborne infection containment.
Article in Journal of the American College of Emergency Physicians open, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 4 citations in OpenAlex.
- A Portable, Foldable Negative-Pressure Aerosol-Containment System (FNPACS) for Aerosol Control During Aerosol-Generating Procedures.Bioengineering (Basel, Switzerland) · 2026Article
- Infection control for COVID-19 in hospital examination room.Scientific reports · 2022Article
- Development and efficacy testing of a portable negative pressure enclosure for airborne infection containment.Journal of the American College of Emergency Physicians open · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesTo overcome the shortage of personal protective equipment and airborne infection isolation rooms (AIIRs) in the COVID-19 pandemic, a collaborative team of research engineers and clinical physicians worked to build a novel negative pressure environment in the hopes of improving healthcare worker and patient safety. The team then sought to test the device's efficacy in generating and maintaining negative pressure. The goal proved prescient as the US Food and Drug Administration (FDA) later recommended that all barrier devices use negative pressure.
methodsInitially, engineers observed simulations of various aerosol- and droplet-generating procedures using hospital beds and stretchers to determine the optimal working dimensions of the containment device. Several prototypes were made based on these dimensions which were combined with filters and various flow-generating devices. Then, the airflow generated and the pressure differential within the device during simulated patient care were measured, specifically assessing its ability to create a negative pressure environment consistent with standards published by the Centers for Disease Control and Prevention (CDC).
resultsThe portable fans were unable to generate any airflow and were dropped from further testing. The vacuums tested were all able to generate a negative pressure environment with the magnitude of pressure differential increasing with the vacuum horsepower. Only the 3.5-horsepower Shop-Vac, however, generated a -3.0 pascal (Pa) pressure gradient, exceeding the CDC-recommended minimum of -2.5 Pa for AIIRs.
conclusionA collaborative team of physicians and engineers demonstrated the efficacy of a prototype portable negative pressure environment, surpassing the negative pressure differential recommended by the CDC.