A method for time-independent film dosimetry: Can we obtain accurate patient-specific QA results at any time postirradiation?
Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed.
- Commissioning of a near real-time in vivo film dosimetry system.Journal of applied clinical medical physics · 2026Article
- On temporal evolution of the response of EBT4 radiochromic films.Journal of applied clinical medical physics · 2026Article
- A review on radiochromic film dosimetry in radiation therapy.Journal of applied clinical medical physics · 2025Review
- Adaptive calibration of Gafchromic EBT3 film using generalized additive neural networks.Scientific reports · 2025Article
- A comparison of three-film analysis software for stereotactic radiotherapy patient-specific quality assurance.Journal of applied clinical medical physics · 2024Article
- A method for time-independent film dosimetry: Can we obtain accurate patient-specific QA results at any time postirradiation?Journal of applied clinical medical physics · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
aimThe aim of this work was twofold. (1) To investigate and present a comparison between EBT3 and EBT-XD in terms of postirradiation color changes. (2) Create an automated workflow to allow radiochromic film (EBT3/XD) to be scanned and converted to dose accurately at any postirradiation time. MATERIALS AND
methodsTen GafChromic EBT-XD calibration films were exposed in 2 Gy increments up to 18 Gy. Calibrates were then scanned at 5-min intervals postirradiation over 24 h using an AutoHotKey script, resulting in 288 TIFF images. Following the 24-h scanning period, a MATLAB script was used to automatically read the tiff images and create a series of 288 calibration curves distinct in time which is termed as the "Temporal Calibration Model" (TCM). The model is saved as a series of polynomial fit coefficients to net optical density as a function of dose, timestamped in 5-min increments. Ten patient-specific film measurements (5 × EBT-XD and 5 × EBT3) were then carried out and scanned using the same 5-min scan intervals from 5 min postirradiation to 24 h postirradiation. The TCM was then automatically applied using eFilmQA software to convert the patient-specific QA films to dose by applying the relevant calibration curve from the TCM, corresponding to the arbitrary postirradiation time that the film was scanned. Each dose plane at postirradiation scan intervals of 5 min up to 20 h was then compared to the ground-truth dose plane using gamma analysis.
resultsGamma pass rates using the TCM at time t, normalized to the pass rate after 20 h postirradiation, were found to have a maximum coefficient of variation of 3% over any postirradiation time. Conversely, not using the TCM resulted in coefficients of variation of up to 39%. Clinical implementation of this method showed an average accuracy of 2.8% when comparing the clinical result to the TCM result.
conclusionsWe have developed a methodology that allows radiochromic film to be accurately used as a dosimeter at any arbitrary scan postirradiation time, whereas previously, waiting periods of 16-24 h before readout were needed to ensure the postirradiation changes had stabilized. The creation of a TCM can enable results from radiochromic film measurements to be obtained quickly postirradiation. Using a conventional single calibration curve generated at 20 h postirradiation can result in gamma pass-rate difference of up to 75% for measurement films scanned at a much shorter postirradiation time.