Trial reportThe American journal of emergency medicine2022
Comparison of Vie Scope® and Macintosh laryngoscopes for intubation during resuscitation by paramedics wearing personal protective equipment.
Trial report in The American journal of emergency medicine, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT04365608 (Safety and Efficacy of Endotracheal Intubation by Paramedics in Suspected/Confirmed COVID-19 Patients Under Cardiac Arrest Using Vie Scope Laryngoscope), which is not on this map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Safety and Efficacy of Endotracheal Intubation by Paramedics in Suspected/Confirmed COVID-19 Patients Under Cardiac Arrest Using Vie Scope Laryngoscope
Who cites it
9 citing papers in PubMed, 23 citations in OpenAlex.
- Vie Scope® versus videolaryngoscopy in expected difficult airways: a randomized controlled trial.Canadian journal of anaesthesia = Journal canadien d'anesthesie · 2023Trial
- Impact of VieScope® on first-attempt success during simulated COVID-19 patients intubation: A randomized cross-over simulation trial.Ulusal travma ve acil cerrahi dergisi = Turkish journal of trauma & emergency surgery : TJTES · 2023Trial
- Performance of the Bebé VieScope Versus Direct Laryngoscopy During Pediatric Cardiopulmonary Resuscitation: A Prospective Randomized Simulation Study.Children (Basel, Switzerland) · 2026Article
- Article
- New Devices, Innovative Technologies, and Non-Standard Techniques for Airway Management: A Narrative Review.Healthcare (Basel, Switzerland) · 2023Review
- Assessment of the Possibility of Using the Laryngoscopes Macintosh, McCoy, Miller, Intubrite, VieScope and I-View for Intubation in Simulated Out-of-Hospital Conditions by People without Clinical Experience: A Randomized Crossover Manikin Study.Healthcare (Basel, Switzerland) · 2023Article
- A Prospective Comparative Study of Laryngeal Glottic View Using the Vie ScopeTherapeutics and clinical risk management · 2023Article
- Airway management in cardiac arrest and outcomes. Author's reply.Intensive care medicine · 2022Article
- A Comparison of the bébé VieScope™ and Direct Laryngoscope for Use While Wearing PPE-AGP: A Randomized Crossover Simulation Trial.Children (Basel, Switzerland) · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors at 5 institutions in 3 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundEndotracheal intubation (ETI) is still the gold standard of airway management, but in cases of sudden cardiac arrest in patients with suspected SARS-CoV-2 infection, ETI is associated with risks for both the patient and the medical personnel. We hypothesized that the Vie Scope® is more useful for endotracheal intubation of suspected or confirmed COVID-19 cardiac arrest patients than the conventional laryngoscope with Macintosh blade when operators are wearing personal protective equipment (PPE).
methodsStudy was designed as a prospective, multicenter, randomized clinical trial performed by Emergency Medical Services in Poland. Patients with suspected or confirmed COVID-19 diagnosis who needed cardiopulmonary resuscitation in prehospital setting were included. Patients under 18 years old or with criteria predictive of impossible intubation under direct laryngoscopy, were excluded. Patients were randomly allocated 1:1 to Vie Scope® versus direct laryngoscopy with a Macintosh blade. Study groups were compared on success of intubation attempts, time to intubation, glottis visualization and number of optimization maneuvers.
resultsWe enrolled 90 out-of-hospital cardiac arrest (OHCA) patients, aged 43-92 years. Compared to the VieScope® laryngoscope, use of the Macintosh laryngoscope required longer times for tracheal intubation with an estimated mean difference of -48 s (95%CI confidence interval [CI], -60.23, -35.77; p < 0.001). Moreover VieScope® improved first attempt success rate, 93.3% vs. 51.1% respectively (odds ratio [OR] = 13.39; 95%CI: 3.62, 49.58; p < 0.001).
conclusionsThe use of the Vie Scope® laryngoscope in OHCA patients improved the first attempt success rate, and reduced intubation time compared to Macintosh laryngoscope in paramedics wearing PPE for against aerosol generating procedures.
trial registrationClinicalTrials registration number NCT04365608.