Infrared Thermography for Measuring Elevated Body Temperature: Clinical Accuracy, Calibration, and Evaluation.

Quanzeng Wang, Yangling Zhou, Pejman Ghassemi, David McBride, Jon P Casamento, T Joshua Pfefer

Open access · goldAbstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 17 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
17citing papers in PubMed, 1 pooled it
4.4field-weighted citation impact, top 5% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

17 citing papers in PubMed, 1 synthesis or guideline pooled it, 41 citations in OpenAlex.

  1. Pooled it
  2. Article
  3. Article
  4. Review
  5. Observational
  6. Article
  7. Review
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Article
  15. Review
  16. Article
  17. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 2 institutions in 1 country.

Quanzeng WangCenter for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA.ORCID 0000-0002-0557-7058
Yangling ZhouCenter for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA.
Pejman GhassemiCenter for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA.ORCID 0000-0003-4810-9454
David McBrideUniversity Health Center, University of Maryland, College Park, MD 20742, USA.
Jon P CasamentoCenter for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA.
T Joshua PfeferCenter for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA.
Center for Devices and Radiological Health · USUniversity of Maryland, College Park · US

Funding

United States Food and Drug Administration 16ECDRH407
6 · The paper itself

Abstract

Infrared thermographs (IRTs) implemented according to standardized best practices have shown strong potential for detecting elevated body temperatures (EBT), which may be useful in clinical settings and during infectious disease epidemics. However, optimal IRT calibration methods have not been established and the clinical performance of these devices relative to the more common non-contact infrared thermometers (NCITs) remains unclear. In addition to confirming the findings of our preliminary analysis of clinical study results, the primary intent of this study was to compare methods for IRT calibration and identify best practices for assessing the performance of IRTs intended to detect EBT. A key secondary aim was to compare IRT clinical accuracy to that of NCITs. We performed a clinical thermographic imaging study of more than 1000 subjects, acquiring temperature data from several facial locations that, along with reference oral temperatures, were used to calibrate two IRT systems based on seven different regression methods. Oral temperatures imputed from facial data were used to evaluate IRT clinical accuracy based on metrics such as clinical bias (Δcb), repeatability, root-mean-square difference, and sensitivity/specificity. We proposed several calibration approaches designed to account for the non-uniform data density across the temperature range and a constant offset approach tended to show better ability to detect EBT. As in our prior study, inner canthi or full-face maximum temperatures provided the highest clinical accuracy. With an optimal calibration approach, these methods achieved a Δcb between ±0.03 °C with standard deviation (σΔcb) less than 0.3 °C, and sensitivity/specificity between 84% and 94%. Results of forehead-center measurements with NCITs or IRTs indicated reduced performance. An analysis of the complete clinical data set confirms the essential findings of our preliminary evaluation, with minor differences. Our findings provide novel insights into methods and metrics for the clinical accuracy assessment of IRTs. Furthermore, our results indicate that calibration approaches providing the highest clinical accuracy in the 37-38.5 °C range may be most effective for measuring EBT. While device performance depends on many factors, IRTs can provide superior performance to NCITs.

Indexed as

Body TemperatureThermographyCalibrationFeverHumansInfrared RaysThermometersclinical accuracyelevated body temperaturefever screeninginfrared thermography

Identifiers

PMID35009758
PMCPMC8749858
OpenAlexW4200395617

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.